7. Error Messages and Troubleshooting
appears or a
measurement
cannot be
complete after
the arm cuff
inflates.
You move or talk during
a measurement and the
arm cuff does not inflate
sufficiently.
Remain still and do not
talk during a
measurement. If “E2”
appears repeatedly,
inflate the arm cuff
manually until the systolic
pressure is 30 to
40 mmHg above your
previous readings. Refer
The systolic pressure is
above 210 mmHg and a
measurement cannot be
taken.
Inflate the arm cuff
manually until the systolic
pressure is 30 to
40 mmHg above your
previous readings. Refer
appears
The arm cuff is inflated
exceeding the maximum
allowable pressure.
Do not touch the arm cuff
and/or bend the air tube
while taking a
measurement.
If inflating the arm cuff
manually, refer to the end
appears
You move or talk during
a measurement.
Vibrations disrupt a
measurement.
Remain still and do not
talk during a
measurement.
Display/
Problem
Possible Cause
Solution