![Omron BP742CANN Instruction Manual Download Page 43](http://html1.mh-extra.com/html/omron/bp742cann/bp742cann_instruction-manual_741509043.webp)
43
9. GUIDANCE AND MANUFACTURER’S DECLARATION
Note1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones
and land mobile radio, AM and FM radio broadcast, and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in
which the OMRON BP742CANN including AC-adapter is used exceeds the applicable RF
compliance level above, the OMRON BP742CANN including AC-adapter should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the OMRON BP742CANN including AC-adapter.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.