background image

SAFETY INFORMATION

INTENDED USE

This product is designed to measure the blood pressure and pulse rate of people within 

the range of the designated arm cuff, following the instructions in this instruction 

manual. It is mainly designed for general household use. Please read the “Safety 

Information” in this instruction manual before using the unit.

To assure the correct use of the product, basic safety measures should always be 

followed including the warnings and cautions listed in this instruction manual.

OPERATING THE DEVICE

Contact your physician for specific information about your blood pressure. 

Self-diagnosis and treatment using measured results may be dangerous. Follow the 

instructions of your physician or licensed healthcare provider.
DO NOT adjust medication based on measurement results from this blood pressure 

monitor. Take medication as prescribed by your physician. Only a physician is 

qualified to diagnose and treat High Blood Pressure.
The monitor is not intended to be a diagnostic device.
If battery fluid should get in your eyes, immediately rinse with plenty of clean 

water. Contact a physician immediately.
Read all of the information in the instruction manual and any other literature in the 

box before operating the unit.
Do not use this device on infants or persons who cannot express their intentions.
Read the Special Conditions section of the instruction manual if your systolic 

pressure is known to be more than 220 mmHg. Inflating to a higher pressure than 

necessary may result in bruising where the cuff is applied.
Operate the device only as intended. Do not use the device for any other purpose.
Dispose of the device, components and optional accessories according to applicable 

local regulations. Unlawful disposal may cause environmental pollution.
Do not use a cellular phone near the device. It may result in an operational failure.
Use only Omron authorized parts and accessories. Parts and accessories not 

approved for use with the device may damage the unit.
Do not use other types of batteries. This may damage the unit.
Do not use in a location with moisture, or a location where water may splash on the 

unit.

CARE AND MAINTENANCE

Do not subject the monitor to strong shocks, such as dropping the unit on the floor.
Do not submerge the device or any of the components in water.
Store the device and the components in a clean, safe location.
Changes or modification not approved by the manufacturer will void the user 

warranty. Do not disassemble or attempt to repair the unit or components.

SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL

WARNING

Indicates a potentially hazardous situation which, if not avoided, 

could result in death or serious injury.

CAUTION

Indicates a potentially hazardous situation which, if not avoided, 

may result in minor or moderate injury to the user or patient or 

damage to the equipment or other property.

BEFORE TAKING A MEASUREMENT

To ensure a reliable reading follow these recommendations:

1.  Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 minutes 

before taking a measurement. Rest for at least 15 minutes before taking the 

measurement.

2.  Stress raises blood pressure. Avoid taking measurements during stressful times.

3.  Measurements should be taken in a quiet place.

4.  Remove tight-fitting clothing from your left arm.

5.  Sit on a chair with your feet flat on the floor. Rest your left arm on table so that the 

cuff is at the same level as your heart.

6.  Remain still and do not talk during the measurement.

7.  Wait 2-3 minutes between measurements. The wait time allows the arteries to return 

to the condition prior to taking the blood pressure measurement. You may need to 

increase the wait time depending on your individual physiological characteristics.

8.  Keep a record of your blood pressure and pulse readings for your physician. A 

single measurement does not provide an accurate indication of your true blood 

pressure. You need to take and record several readings over a period of time. Try to 

measure your blood pressure at the same time each day for consistency.

INSTRUCTION MANUAL

5 series

Blood Pressure Monitor

Model  BP742

INTRODUCTION

Thank you for purchasing the OMRON

®

 BP742 5 series Automatic Blood Pressure 

Monitor.

Your new blood pressure monitor uses the oscillometric method of blood pressure 

measurement. This means the monitor detects your blood’s movement through your 

brachial artery and converts the movements into a digital reading. An oscillometric 

monitor does not need a stethoscope so the monitor is simple to use.

The BP742 comes with the following components:

• Monitor • Wide Range Cuff  • 4 “AA” Alkaline Batteries

• Instruction Manual printed in English and Spanish • Quick Start Guide

The BP742 5 series Automatic Blood Pressure Monitor is intended for home use.

Please read this instruction manual thoroughly before using the unit.

Please keep for future reference.

For specific information about your own blood pressure, CONSULT YOUR 

DOCTOR.

Fill in for future reference.

DATE PURCHASED: __________________

SERIAL NUMBER: ____________________

• Staple your purchase receipt here
• Register your product on-line at
  www.register-omron.com

SAVE THESE INSTRUCTIONS

LIMITED WARRANTY

Your BP742 5 series Automatic Blood Pressure Monitor, excluding the arm cuff and 

batteries, is warranted to be free from defects in materials and workmanship appearing 

within 2 years from the date of purchase, when used in accordance with the instructions 

provided with the monitor. The arm cuff is warranted to be free from defects in 

materials and workmanship appearing within one year from the date of purchase when 

the monitor is used in accordance with the instructions provided with the monitor. The 

above warranties extend only to the original retail purchaser.
We will, at our option, repair or replace without charge any monitor or arm cuff 

covered by the above warranties. Repair or replacement is our only responsibility and 

your only remedy under the above warranties.
To obtain warranty service contact Customer Service by calling 

1-800-634-4350

 for the 

address of the repair location and the return shipping and handling fee.
Enclose the Proof of Purchase. Include a letter, with your name, address, phone 

number, and description of the specific problem. Pack the product carefully to prevent 

damage in transit. Because of possible loss in transit, we recommend insuring the 

product with return receipt requested.

THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN 

CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS 

ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING IMPLIED 

WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR 

PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS THAT MAY BE 

IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION TO THE PERIOD 

OF THE ABOVE EXPRESS WARRANTY.

OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER 

SPECIAL, INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS, 

EXPENSES OR DAMAGES.

This warranty provides you with specific legal rights, and you may have other rights 

that vary by jurisdiction. Because of special local requirements, some of the above 

limitations and exclusions may not apply to you.

FOR CUSTOMER SERVICE

Visit our web site at:                                                              www.omronhealthcare.com

Call toll free:                                                                                            1-800-634-4350

Note:

POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only)
This product has been tested and found to comply with the limits for a Class B digital 

device, pursuant to part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful interference 

in a residential installation. The product generates, uses, and can radiate radio 

frequency energy and, if not installed and used in accordance with the instructions, 

may cause harmful interference to radio communications. However, there is no 

guarantee that interference will not occur in a particular installation. If the product does 

cause harmful interference to radio or television reception, which can be determined by 

turning the product on and off, the user is encouraged to try to correct the interference 

by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the product and the receiver.
• Connect the product into an outlet on a circuit different from that to which the 

receiver is connected.

• Consult the dealer or an experienced radio/TV technician for help.

POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only)
This digital apparatus does not exceed the Class B limits for radio noise emissions 

from digital apparatus as set out in the interference-causing equipment standard entitled 

“Digital Apparatus”, ICES-003 of the Canadian Department of Communications.
Cet appareil numérique respecte les limites de bruits radioeléctriques applicables aux 

appareils numériques de Clase B prescrites dans la norme sur le materiel brouilleur: 

“Appareils Numériques”, ICES-003 édictée par le minister des communications.

Changes or modifications not expressly approved by the party responsible for 

compliance could void the user’s authority to operate the equipment.

SPECIFICATIONS

Model:
Display:
Measurement Range:
Accuracy/Calibration:

Inflation:
Deflation:
Rapid Air Release:
Measurement Method:
Power Source:
Battery Life:
Operating Temperature 
/Humidity:
Storage Temperature 
/Humidity/Air Pressure:
Main Unit Weight:

Main Unit Dimensions:

Cuff Size:

Cuff Circumference:

Memory:
Contents:

Optional Accessories:

BP742 (HEM-7200-Z)
LCD Digital Display
Pressure: 0 to 299 mmHg, Pulse: 40 to 180/min.
Pressure: ±3mmHg or 2% of reading

Pulse: ±5% of reading
Automatic by electric pump
Automatic pressure release valve
Automatic exhaust valve
Oscillometric method
1.5V 4 “AA” batteries
Approx. 300 uses with 4 new alkaline batteries
50ºF to 104ºF (10ºC to 40ºC) /15 to 90% RH

-4ºF to 140ºF (-20ºC to 60ºC) /10 to 95% RH

700-1060 hPa
Approximately 12 oz. (340g)

not including batteries
Approx. 5 l/2"(l) × 4 7/8"(w) × 3 3/8" (h)

(141 mm × 123 mm × 85 mm)
Approx. 6"(w) × 22 1/8"(l) (151 mm × 563 mm)

Cuff tube 35 3/8" (900 mm)
Fits arm circumferences 9" to 17"

(220 mm to 420 mm)
Up to 30 readings per user
Main Unit, Wide Range Cuff, 4 “AA” Alkaline Batteries, 

Instruction Manual, and Quick Start Guide
AC Adapter

NOTE: These specifications are subject to change without notice.

Manufactured for OMRON HEALTHCARE Co., Ltd.

Distributed by OMRON HEALTHCARE, INC.

1925 West Field Court

Lake Forest, IL 60045 U.S.A.

www.omronhealthcare.com

© 2012 OMRON HEALTHCARE, INC.
Made in Vietnam

FCC STATEMENT

HEM-7200-Z(V)_C_M02_120928.pdf

SETTING THE DATE AND TIME

Set the monitor to the current date and time before taking a measurement for the first 

time. If the date and time are not set correctly, the measurement values stored in the 

memory and the average values may not be correct.

1.  TO START THE SETTING

Press the Date/Time Setting button (     ) to adjust the date and time.

The year flashes on the display.

2.  SETTING THE YEAR

The year can be set between 2010 and 2040. When the 

display reaches 2040, it will return to 2010.

Press the Up button (      ) to advance by one year.

Press the Date/Time Setting button (     ) to set the current 

year.

The month flashes on the display.

NOTES: 

 

Press the Down button (      ) to decrease the 

date and time values.

• Press and hold the button to increase or 

decrease the date and time values faster.

3.  SETTING THE MONTH

Press the Up button (      ) to advance by one month.

Press the Date/Time Setting button (     ) to set the current 

month.

The day flashes on the display.

4.  SETTING THE DAY

Press the Up button (      ) to advance by one day.

Press the Date/Time Setting button (     ) to set the current 

day.

The hour flashes on the display.

5.  SETTING THE HOUR

The time is set using AM or PM.

Press the Up button (      ) to advance by one hour.

Press the Date/Time Setting button (     ) to set the current

hour.

The minute flashes on the display.

6.  SETTING THE MINUTE

Press the Up button (      ) to advance by one minute.

Press the Date/Time Setting button (     ) to set the minute setting.

7.  TO TURN THE MONITOR OFF

Press the START/STOP button.

NOTE:

 The monitor will automatically turn off five minutes after the minute has 

been set.

WARNING

If battery fluid should get in your eyes, immediately rinse with

plenty of clean water. Contact a physician immediately.

CAUTION

Read all of the information in the instruction manual and any other

literature in the box before operating the unit.

CAUTION

This device is intended for use in measuring blood pressure and

pulse rate in the adult population. Do not use this device on infants

or persons who cannot express their intentions.

APPLYING THE CUFF ON THE LEFT ARM

1.  Make sure the air plug is securely inserted 

in the main unit.

2.  Remove tight-fitting clothing from your upper 

arm.

3.  Apply the cuff to your left upper arm so 

the arrow is centered on the inside of your arm 

and aligned with your middle finger. 

The air tube runs down the inside of your arm. 

The bottom of the cuff should be 

approximately 1/2 " above your elbow.

4. Wrap the cuff firmly in place around your 

arm using the cloth fastening.

APPLYING THE CUFF ON THE RIGHT ARM

When taking a measurement using the right arm, use this instruction for Step 3.

Apply the cuff to your right upper arm so 

the arrow and the air tube runs along the inside 

of your arm. The bottom of the cuff should be 

approximately 1/2 " above your elbow.

NOTE: 

Be careful not to rest your arm on the air tube. This will restrict the flow of air 

to the cuff.

APPLYING THE ARM CUFF

1/2 inch (1 - 2 cm)

Air Tube

Arrow

Left Arm

Right Arm

Arrow

TAKING A MEASUREMENT

1.  Press the START/STOP button to turn the monitor on.

All symbols appear on the display.

2.  Select your USER ID (A or B).

3.  Sit on a chair with your feet flat on the floor. 

Place your arm on a table so the cuff is level 

with your heart. 

Keep still and do not talk during measurement.

4.  Press the START/STOP button.

The cuff starts to inflate automatically. As the cuff inflates, the monitor automatically 

determines your ideal inflation level. 

Remain still and do not move your arm until the entire measurement process is completed.

NOTE: 

To stop the inflation or measurement, push the START/STOP button. The 

monitor will stop inflating, start deflating, and will turn off.

5.  Inflation stops automatically and the measurement is started.

As the cuff deflates, decreasing numbers appear on the display and the Heartbeat 

Symbol (     ) flashes.

6.  When the measurement is complete, the arm cuff completely 

deflates. Your blood pressure and pulse rate are displayed.

7.  Press the START/STOP button to turn the monitor off.

NOTE: 

The monitor will automatically turn off after 

five minutes.

CARE AND MAINTENANCE

To keep your blood pressure monitor in the best condition and protect the unit from 

damage, follow the directions listed below:
Do not forcefully

 bend the arm cuff or air tube. Do not fold tightly.

Clean the monitor with a soft dry cloth.

Do not use any abrasive or volatile cleaners.

Do not attempt to clean the cuff.

CAUTION

Do not submerge the device or any of the components in water. Do

not subject the monitor to extreme hot or cold temperatures,

humidity or direct sunlight.

CAUTION

Store the device and the components in a clean, safe location.

CAUTION

Do not subject the monitor to strong shocks, such as dropping the unit on the floor.

Remove the batteries

 if the unit will not be used for three months or longer. Always 

replace all the batteries with new ones at the same time.

Use the unit in a manner consistent with the instructions provided in this manual.

CAUTION

Use only Omron authorized parts and accessories. Parts and accessories 

not approved for use with the device may damage the unit.

CAUTION

Changes or modification not approved by the manufacturer will void the user 

warranty. Do not disassemble or attempt to repair the unit or components.

TROUBLESHOOTING TIPS

PROBLEM

No power.

No display appears on the unit.

Measurement values appear too 

high or too low.

CAUSES AND SOLUTIONS

Replace all four batteries with new ones.

Check the battery installation for proper placement 

of the battery polarities.
Blood pressure varies constantly. Many factors 

including stress, time of day, how you wrap the 

cuff, may affect your blood pressure. 

Review the sections “Before Taking a 

Measurement” and “Taking a Measurement”.

ERROR INDICATORS

SYMBOL

CAUSE

Device error.

CORRECTION

Keep still and take another 

measurement.

Replace the four batteries. 

Refer to “Battery Installation”.

Contact Customer Service.

Remove the arm cuff. Wait 2-3 

minutes and then take another 

measurement. Repeat the steps in 

section “Applying the Arm Cuff”. 

If this error continues to appear, 

contact your doctor. 

Batteries are low.

Irregular or weak

pulses are detected.

Movement during 

measurement.

Monitor could not 

detect pulse wave.

Cuff over-inflated.

Cuff under-inflated.

Cuff not applied 

correctly.

Take another measurement and 

remain still until the measurement 

is complete.

Remove the arm cuff.

Read “Taking a Measurement”.

Wait 2-3 minutes.

Take another measurement.

1.  Press the     indicator on the battery cover and 

slide the cover off in the direction of the arrow.

2.  Install 4 “AA” size batteries so the + (positive) and

 - (negative) polarities match the polarities of 

the battery compartment as indicated.

3.  Replace the battery cover.

BATTERY REPLACEMENT

Low Battery Indicator

When the Low Battery Indicator appears on the display, turn the monitor off and 

remove all the batteries. Replace with four new batteries at the same time. Long-life 

alkaline batteries are recommended.

NOTES:  

 

Turn the unit off before replacing the batteries. If the batteries are removed 

while the unit is still on, the date and time will be reset to that of the 

previous use. However, the measurement results are not deleted.

• When the batteries are replaced, you may need to reset the date and time. If 

“-:--” appears on the display, refer to “Setting the Date and Time”.

• The supplied batteries may have a shorter life.

CAUTION

Dispose of the device, components and optional accessories according to 

applicable local regulations. Unlawful disposal may cause environmental 

pollution.

BATTERY INSTALLATION

UNIT DISPLAY

Systolic blood pressure

Diastolic blood pressure

Memory symbol

(Displayed when viewing
values stored in memory)

Average Value symbol

Heartbeat symbol /
Hypertension Indicator

Irregular Heartbeat symbol

User ID symbol (A/B)

Pulse display
Deflation symbol

Battery low indicator

Date/Time Display

Movement error symbol

KNOW YOUR UNIT

Main Unit:

Components:

START/STOP button

USER ID 
Selection Switch
Up/Down (                    ) buttons

Air Jack

Date/Time setting (      ) 
button

Wide Range Cuff

Air tube

Air plug

Quick Start 
Guide

Instruction 
Manual

Display

INSTRUCTION MANUAL

INTRODUCTION

Thank you for purchasing the OMRON® BP742 5 series Automatic Blood Pressure 

Monitor.

Your new digital blood pressure monitor uses the oscillometric method of blood 

pressure measurement. This means the monitor detects your blood’s movement through 

your brachial artery and converts the movements into a digital reading. An 

oscillometric monitor does not need a stethoscope so the monitor is simple to use.

The BP742 comes with the following components:

• Monitor • Standard Adult Arm Cuff 

• Instruction Manual printed in English and Spanish • Quick Start Guide

The BP742 IntelliSense® Automatic Inflation Blood Pressure Monitor is intended for 

home use.

Fill in for future reference.

DATE PURCHASED: __________________

SERIAL NUMBER: ____________________

Staple

 

y

our 

pu

r

chase

 rec

eipt

 

here

SAVE THESE INSTRUCTIONS

5329898-7C

Battery compartment

AC adapter Jack

BP742

5 Series Arm Blood Pressure

Monitor

5 series

Blood Pressure Monitor

Model  BP742

DISPLAY SYMBOLS

IRREGULAR HEARTBEAT SYMBOL (       )

When the monitor detects an irregular rhythm two or more times during the 

measurement, the Irregular Heartbeat Symbol (       ) will appear on the display with the 

measurement values.
An irregular heartbeat rhythm is defined as 

a rhythm that varies by less than 25% from 

the average rhythm or 25% more than the 

average rhythm detected while the monitor 

is measuring the systolic and diastolic 

blood pressure.
If the irregular heartbeat symbol (        ) 

displays with your measurement results, 

we recommend you consult your 

physician. Follow the directions of your 

doctor.

MOVEMENT ERROR SYMBOL (       )

The Movement Error Symbol (        ) is displayed if you move your body during the 

measurement. Please remove the arm cuff, and wait 2-3 minutes. Reapply the arm cuff 

and take another measurement.

AVERAGE VALUE SYMBOL (       )

The Average Value Symbol (        ) is displayed when you press the memory button. 

The most recent average reading appears on the display screen.

HEARTBEAT SYMBOL/HYPERTENSION INDICATOR (      )

TAKING A MEASUREMENT

The Heartbeat Symbol (      ) flashes on the display during the measurement. The 

Heartbeat Symbol (      ) flashes at every heartbeat.

When the measurement is complete, the Heartbeat Symbol (      ) flashes on the 

display with your blood pressure and pulse rate if the reading is above 135 for the 

Systolic Blood Pressure value and/or above 85 for the Diastolic Blood Pressure 

value.

USING THE MEMORY FUNCTION

The Heartbeat Symbol (      ) flashes when the blood pressure reading stored in the 

memory is above 135 for the Systolic Blood Pressure value and/or above 85 for the 

Diastolic Blood Pressure value.

Internationally recognized guidelines for upper limit of normal home blood pressure 

readings are as follows:

This is a general guideline because blood pressure depends on age and morbidity.

A lower home blood pressure goal may be recommended for certain patients, 

including diabetic patients, pregnant women, and patients with renal failure.

OMRON HEALTHCARE recommends you contact your physician if the Heartbeat 

Symbol (      ) flashes on the display with your measurement values.

WARNING

The monitor is not intended to be a diagnostic device.

WARNING

Contact your physician for specific information about your blood pressure. 

Self-diagnosis and treatment using measured results may be dangerous. 

Follow the instructions of your physician or licensed healthcare provider.

Normal Heartbeat

Irregular Heartbeat

Short Long

Blood 

pressure

Pulse

Pulse

Blood 

pressure

Home Guideline for Upper Limit of Normal BP

Systolic Blood Pressure

135 mmHg

Diastolic Blood Pressure

85 mmHg

SPECIAL CONDITIONS

CAUTION

Inflating to a higher pressure than necessary may result in bruising where 

the cuff is applied.

If your systolic pressure is known to be more than 220 mmHg, press and 

hold the START/STOP button until the monitor inflates 30 to 40 mmHg 

higher than your expected systolic pressure.

NOTES:  

 

Do not apply more pressure than necessary.

• The monitor will not inflate above 299 mmHg.

When the measurement is complete, the arm cuff completely deflates.

Your blood pressure and pulse rate are displayed.

USING THE MEMORY FUNCTION

The monitor is designed to store the blood pressure and the pulse rate in the memory 

for two people (USER A and USER B) every time a measurement is completed.

The monitor automatically stores up to 30 sets of measurement values (blood pressure 

and pulse rate) for each user (A and B). When 30 sets of measurement values are 

stored, the oldest record is deleted to save the most recent values.

AVERAGING FUNCTION

The monitor calculates the average reading based on the three most recent sets of 

measurement values taken within 10 minutes of the most recent reading.

NOTE:   

If two sets of measurement values are stored in the memory for the 10 

minutes period, the average is based  on the two sets of measurement values. 

If one set of measurement values is stored, this is displayed as the average.

TO DISPLAY THE MEASUREMENT VALUES

1.  Press the START/STOP button to turn the monitor on.

2.  Select your USER ID (A or B).

3.  Press the Up or Down button (                ). The Average Value Symbol (      ) appears 

on the display when the average reading is shown.

4.  Press the Up or Down button (                ) to display the sets of measurement values.

Press the Down button (       ) repeatedly

Average value → most recent set of measurement values → the older readings → 

back to the average value

Press the Up button (       ) repeatedly

Average value → oldest set of measurement values → the more recent readings → 

back to the average value

NOTES:  

 

The measurement value number is displayed 

briefly in the Pulse Display area when you press 

the Up/Down buttons (                ).

• The date and time are alternately displayed with 

the measurement values.

• If the date and time are not set, “-:--” is displayed 

with the measurement values.

• Press the button repeatedly to display the next 

values.

• Press and hold the button to display the values 

faster.

TO DELETE ALL VALUES STORED IN THE MEMORY

The values stored in the memory are deleted by USER ID.

You cannot partially delete values stored in the memory. All values for

the user you select will be deleted.

1. Press the START/STOP button to turn the monitor on.

2. Select your USER ID (A or B).

3. While holding the Down button (      ), press the START/STOP button down for 

more than five seconds until the display appears as it does in the following 

illustration.

4. Press the START/STOP button to turn the monitor off.

NOTE: 

The monitor will automatically turn off after five minutes.

Alternating 
display

Reviews: