TROUBLESHOOTING TIPS

PROBLEM

No power.
No display appears on the unit.

Measurement values appear too 
high or too low.

CAUSES AND SOLUTIONS

Replace all four batteries with new ones.
Check the battery installation for proper placement 
of the battery polarities.
Blood pressure varies constantly. Many factors 
including stress, time of day, how you wrap the 
cuff, may affect your blood pressure. 
Review the sections “Before Taking a 
Measurement” and “Taking a Measurement”.

CARE AND MAINTENANCE

To keep your digital blood pressure monitor in the best condition and protect the unit 
from damage, follow the directions listed below:

Do not forcefully

 bend the arm cuff or air tube. Do not fold tightly.

Clean the monitor with a soft dry cloth.

Do not use any abrasive or volatile cleaners.
Do not attempt to clean the cuff.

CAUTION

Do not submerge the device or any of the components in water. Do
not subject the monitor to extreme hot or cold temperatures,
humidity or direct sunlight.

CAUTION

Store the device and the components in a clean, safe location.

CAUTION

Do not subject the monitor to strong shocks, such as dropping the unit on the floor.

Remove the batteries

 if the unit will not be used for three months or longer. Always 

replace all the batteries with new ones at the same time.

Use the unit consistent with the instruction provided in this manual.

Use only authorized parts and accessories.

Use the unit in a manner consistent with the instructions provided in this manual.

CAUTION

Use only Omron authorized parts and accessories. Parts and accessories 
not approved for use with the device may damage the unit.

CAUTION

Changes or modification not approved by the manufacturer will void the user 
warranty. Do not disassemble or attempt to repair the unit or components.

WARRANTY INFORMATION

LIMITED WARRANTIES

Your BP710 IntelliSense

®

 Automatic Blood Pressure Monitor, excluding the arm cuff, 

is warranted to be free from defects in materials and workmanship appearing within 2 
years from the date of purchase, when used in accordance with the instructions 
provided with the monitor. The arm cuff is warranted to be free from defects in 
materials and workmanship appearing within one year from the date of purchase when 
the monitor is used in accordance with the instructions provided with the monitor. The 
above warranties extend only to the original retail purchaser.
We will, at our option, repair or replace without charge any monitor or arm cuff 
covered by the above warranties. Repair or replacement is our only responsibility and 
your only remedy under the above warranties.
To obtain warranty service contact Customer Service by calling 

1-800-634-4350

 for 

the address of the repair location and the return shipping and handling fee.
Enclose the Proof of Purchase. Include a letter, with your name, address, phone 
number, and description of the specific problem. Pack the product carefully to prevent 
damage in transit. Because of possible loss in transit, we recommend insuring the 
product with return receipt requested.

ALL IMPLIED WARRANTIES, INCLUDING BUT NOT LIMITED TO THE 
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR 
PARTICULAR PURPOSE, ARE LIMITED TO THE DURATION OF THE 
APPLICABLE WRITTEN WARRANTY ABOVE.

 Some states do not allow 

limitations on how long an implied warranty lasts, so the above limitation may not 
apply to you.

OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER 
INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS, EXPENSES OR 
DAMAGES.

 Some states do not allow the exclusion or limitation of incidental or 

consequential damages, so the above exclusions may not apply to you.
This warranty gives you specific legal rights, and you may also have other rights which 
may vary from state to state.

FOR CUSTOMER SERVICE

Visit our web site at:                                                              www.omronhealthcare.com
Call toll free:                                                                                            1-800-634-4350

FCC STATEMENT

Note:

POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only)
This product has been tested and found to comply with the limits for a Class B 
digital device, pursuant to part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful 
interference in a residential installation. The product generates, uses, and can radiate 
radio frequency energy and, if not installed and used in accordance with the 
instructions, may cause harmful interference to radio communications. However, 
there is no guarantee that interference will not occur in a particular installation. If 
the product does cause harmful interference to radio or television reception, which 
can be determined by turning the product on and off, the user is encouraged to try to 
correct the interference by one or more of the following measures:

• Reorient or relocate the receiving antenna

.

• Increase the separation between the product and the receiver

.

• Connect the product into an outlet on a circuit different from that to which the 

receiver is connected.

• Consult the dealer or an experienced radio/TV technician for help

.

POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only)
This digital apparatus does not exceed the Class B limits for radio noise emissions 
from digital apparatus as set out in the interference-causing equipment standard 
entitled “Digital Apparatus”, ICES-003 of the Canadian Department of 
Communications.
Cet appareil numérique respecte les limites de bruits radioeléctriques applicables 
aux appareils numériques de Clase B prescrites dans la norme sur le materiel 
brouilleur: “Appareils Numériques”, ICES-003 édictée par le minister des 
communications.
Changes or modifications not expressly approved by the party responsible for 
compliance could void the user’s authority to operate the equipment.

SAFETY INFORMATION

To assure the correct use of the product, basic safety measures should always be 
followed including the warnings and cautions listed in this instruction manual.

OPERATING THE DEVICE

Contact your physician for specific information about your blood pressure. 
Self-diagnosis and treatment using measured results may be dangerous. Follow the 
instructions of your physician or licensed healthcare provider.
The monitor is not intended to be a diagnostic device.
If battery fluid should get in your eyes, immediately rinse with plenty of clean 
water. Contact a physician immediately.
Read all of the information in the instruction manual and any other literature in the 
box before operating the unit.
This device is intended for use in measuring blood pressure and pulse rate in the 
adult population. Do not use this device on infants or persons who cannot express 
their intentions.
Read the Special Conditions section of the instruction manual if your systolic 
pressure is known to be more than 220 mmHg. Inflating to a higher pressure than 
necessary may result in bruising where the cuff is applied.
Operate the device only as intended. Do not use the device for any other purpose.
Dispose of the device, components and optional accessories according to applicable 
local regulations. Unlawful disposal may cause environmental pollution.
Do not use a cellular phone near the device. It may result in an operational failure.
Use only Omron authorized parts and accessories. Parts and accessories not 
approved for use with the device may damage the unit.
Use only 1.5 V “AAA” alkaline batteries with this device. Do not use other types of 
batteries. This may damage the unit.

CARE AND MAINTENANCE

Do not subject the monitor to strong shocks, such as dropping the unit on the floor.
Do not submerge the device or any of the components in water.
Store the device and the components in a clean, safe location.
Changes or modification not approved by the manufacturer will void the user 
warranty. Do not disassemble or attempt to repair the unit or components.

SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL

WARNING

Indicates a potentially hazardous situation which, if not avoided, 
could result in death or serious injury.

CAUTION

Indicates a potentially hazardous situation which, if not avoided, 
may result in minor or moderate injury to the user or patient or 
damage to the equipment or other property.

BEFORE TAKING A MEASUREMENT

To ensure a reliable reading follow these recommendations:
1.  Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 minutes 

before taking a measurement. Rest for at least 15 minutes before taking the 
measurement.

2.  Stress raises blood pressure. Avoid taking measurements during stressful times.
3.  Measurements should be taken in a quiet place.
4.  Remove tight-fitting clothing from your left arm.
5.  Sit on a chair with your feet flat on the floor. Rest your left arm on table so that the 

cuff is at the same level as your heart.

6.  Remain still and do not talk during the measurement.
7.  Wait 2-3 minutes between measurements. The wait time allows the arteries to return 

to the condition prior to taking the blood pressure measurement. You may need to 
increase the wait time depending on your individual physiological characteristics.

8.  Keep a record of your blood pressure and pulse readings for your physician. A 

single measurement does not provide an accurate indication of your true blood 
pressure. You need to take and record several readings over a period of time. Try to 
measure your blood pressure at the same time each day for consistency.

BATTERY REPLACEMENT

Low Battery Indicator

When the Low Battery Indicator appears on the display, turn the monitor off and 
remove all the batteries. Replace with four new batteries at the same time. Long-life 
alkaline batteries are recommended.

CAUTION

Dispose of the device, components and optional accessories according to 
applicable local regulations. Unlawful disposal may cause environmental 
pollution.

WARNING

If battery fluid should get in your eyes, immediately rinse with
plenty of clean water. Contact a physician immediately.

BATTERY INSTALLATION

CAUTION

Use only 1.5V “AAA” alkaline batteries with this device.

1.  Press the     indicator on the battery cover and 

slide the cover off in the direction of the arrow.

2.  Install 4 “AAA” size batteries so the + (positive) and

 - (negative) polarities match the polarities of 
the battery compartment as indicated.

NOTE: 

4 “AAA” size batteries are not included 
in this unit.

3.  Replace the battery cover.

INSTRUCTION MANUAL

Automatic

Blood Pressure Monitor

Model 

BP710

UNIT DISPLAY

Systolic blood pressure

Diastolic blood pressure

Memory symbol

(Displayed when viewing
values stored in memory)

Heartbeat symbol
Pulse display

Deflation symbol
Battery low indicator

Irregular Heartbeat symbol

USING THE MEMORY FUNCTION

Every time you complete a measurement the monitor stores the blood pressure and pulse 
rate in the memory. The monitor automatically stores up to 14 sets of measurement values 
(systolic pressure, diastolic pressure, and pulse rate). When 14 sets of measurement 
values are stored, the oldest record is deleted to save the most recent values.
1.  Press the Memory (M) button. 

The most recent set of measurement values appears on the display. 
The values are displayed from the most recent to the oldest.

NOTE: 

The number of the reading stored in the memory appears for a second 
before the pulse rate is displayed. The newest set is numbered “1”.

2.  Press the START/STOP button to turn the unit off.

TO DELETE ALL THE VALUES STORED IN MEMORY

You cannot partially delete values stored in the memory. All values will be deleted.
1.  Press the Memory (M) button.

Then while holding it down, press the START/STOP button down for more than 2 
seconds until the display appears as illustrated below.

2.  Press the START/STOP button to turn the unit off.

NOTE: 

The monitor will automatically turn off after five minutes.

SPECIFICATIONS

Model:
Display:
Measurement Range:
Accuracy/Calibration:

Inflation:
Deflation:
Rapid Air Release:
Pressure Detection:
Measurement Method:
Pulse Wave Detection:
Power Source:
Battery Life:
Operating Temperature 
/Humidity:
Storage Temperature 
/Humidity/Air Pressure:
Main Unit Weight:

Main Unit Dimensions:

Cuff Size:

Cuff Circumference:

Memory:
Contents:

Optional Accessories:

BP710 (HEM-7113-Z)
LCD Digital Display
Pressure: 0 to 299 mmHg, Pulse: 40 to 180/min.
Pressure: ±3mmHg or 2% of reading
Pulse: ±5% of reading
Automatic by electric pump
Automatic pressure release valve
Automatic exhaust valve
Capacitive pressure sensor
Oscillometric method
Capacitance type pressure sensor
1.5V 4 “AAA” batteries
Approx. 300 uses with 4 new alkaline batteries
50ºF to 104ºF (10ºC to 40ºC) /15 to 90% RH

-4ºF to 140ºF (-20ºC to 60ºC) /10 to 95% RH
700-1060 hPa
Approximately 8 5/8 oz. (245g)
not including batteries
Approx. 5"(l) × 4 1/8"(w) × 2 1/2" (h)
(128 mm × 104 mm × 64 mm)
Approximately 5 3/4"(w) × 17 1/2"(l) (146 mm × 446 mm)
Cuff tube 23 5/8" (600 mm)
Fits arm circumferences 9" to 13"
(220 mm to 320 mm)
Up to 14 readings
Main Unit, Standard Adult Arm Cuff, Instruction Manual, 
and Quick Start Guide
Large Cuff (13"-17")

NOTE: These specifications are subject to change without notice.

OMRON HEALTHCARE, INC.
1200 Lakeside Drive
Bannockburn, Illinois 60015
www.omronhealthcare.com
Copyright © 2009 Omron Healthcare, Inc.

Made in China

INTRODUCTION

Thank you for purchasing the OMRON

®

 BP710 IntelliSense

®

 Automatic Blood 

Pressure Monitor.

Your new digital blood pressure monitor uses the oscillometric method of blood 
pressure measurement. This means the monitor detects your blood’s movement through 
your brachial artery and converts the movements into a digital reading. An 
oscillometric monitor does not need a stethoscope so the monitor is simple to use.
The BP710 comes with the following components:
• Monitor • Standard Adult Arm Cuff 
• Instruction Manual printed in English and Spanish • Quick Start Guide
The BP710 IntelliSense

®

 Automatic Inflation Blood Pressure Monitor is intended for 

home use.

Fill in for future reference.

DATE PURCHASED: __________________

SERIAL NUMBER: ____________________

Staple your purchase receipt here

SAVE THESE INSTRUCTIONS

KNOW YOUR UNIT

Main Unit:

Components:

Display

START/STOP button

Memory (M) button

Battery compartment

Air jack

Standard Adult Arm Cuff

Air tube

Air plug

DISPLAY SYMBOLS

IRREGULAR HEARTBEAT SYMBOL (      )

When the monitor detects an irregular rhythm two or more times during the 
measurement, the Irregular Heartbeat Symbol (      ) will appear on the display with the 
measurement values.
An irregular heartbeat rhythm is defined 
as a rhythm that is more than 25% 
slower or 25% faster from the average 
rhythm detected while the monitor is 
measuring the systolic and diastolic 
blood pressure.
If the irregular heartbeat symbol (      ) 
displays with your measurement results, 
we recommend you consult your 
physician. Follow the directions of your 
doctor.

HEARTBEAT SYMBOL/HYPERTENSION INDICATOR (      )

TAKING A MEASUREMENT

The Heartbeat Symbol (      ) flashes on the display during the measurement. The 
Heartbeat Symbol (      ) flashes at every heartbeat.
When the measurement is complete, the Heartbeat Symbol (      ) flashes on the 
display with your blood pressure and pulse rate if the reading is above 135 for the 
Systolic Blood Pressure value and/or above 85 for the Diastolic Blood Pressure 
value.

USING THE MEMORY FUNCTION

The Heartbeat Symbol (      ) flashes when the blood pressure reading stored in the 
memory is above 135 for the Systolic Blood Pressure value and/or above 85 for the 
Diastolic Blood Pressure value.

The American Heart Association recommends the following guideline for upper limit 
of normal for home blood pressure readings:

This is a general guideline because blood pressure depends on age and morbidity. 
The AHA recommends a lower home blood pressure goal for certain patients, 
including diabetic patients, pregnant women, and patients with renal failure

1

.

Omron Healthcare recommends you contact your physician if the Heartbeat Symbol 
(      ) flashes on the display with your measurement values.

1

American Heart Association Scientific Statement 2004

WARNING

The monitor is not intended to be a diagnostic device.

WARNING

Contact your physician for specific information about your blood pressure. 
Self-diagnosis and treatment using measured results may be dangerous. 
Follow the instructions of your physician or licensed healthcare provider.

Normal Heartbeat

Irregular Heartbeat

Short Long

Blood 

pressure

Pulse

Pulse

Blood 

pressure

AHA Home Guideline for Upper Limit of Normal BP

Systolic Blood Pressure

135 mmHg

Diastolic Blood Pressure

85 mmHg

INSTRUCTION MANUAL

Automatic

Blood Pressure Monitor

Model 

BP710

INTRODUCTION

Thank you for purchasing the OMRON  BP710 Automatic Blood Pressure Monitor.

Fill in for future reference.

DATE PURCHASED: __________________

SERIAL NUMBER: ____________________

TAKING A MEASUREMENT

1.  Sit on a chair with your feet flat on the floor. 

Place your arm on a table so the cuff is level 
with your heart. 
Keep still and do not talk during measurement.

2.  Press the START/STOP button.

All symbols appear on the display. 
The cuff starts to inflate automatically. As the cuff inflates, the monitor automatically 
determines your ideal inflation level. 
Remain still and do not move your arm until the entire measurement process is completed.

NOTE: 

To stop the inflation or measurement, push the START/STOP button. The 
monitor will stop inflating, start deflating, and will turn off.

3.  Inflation stops automatically and the measurement is started.

As the cuff deflates, decreasing numbers appear on the display and the Heartbeat 
Symbol (     ) flashes.

4.  When the measurement is complete, the arm cuff completely 

deflates. Your blood pressure and pulse rate are displayed.

5.  Press the START/STOP button to turn the monitor off.

NOTE: 

The monitor will automatically turn off after five minutes.

SPECIAL CONDITIONS

CAUTION

Inflating to a higher pressure than necessary may result in bruising where 
the cuff is applied.
If your systolic pressure is known to be more than 220 mmHg, press and 
hold the START/STOP button until the monitor inflates 30 to 40 mmHg 
higher than your expected systolic pressure.

NOTES:  

 

Do not apply more pressure than necessary.

• The monitor will not inflate above 299 mmHg.

When the measurement is complete, the arm cuff completely deflates.
Your blood pressure and pulse rate are displayed.

ERROR INDICATORS

SYMBOL

CAUSE

Monitor could not 
detect pulse wave.

Cuff over-inflated.

Batteries are worn.

Device error.

Cuff under-inflated.
Cuff not applied 
correctly.

CORRECTION

Take another measurement and 
remain still until the measurement 
is complete.

Replace the four batteries. Refer to 
“Battery Installation”.

Contact Customer Service.

Remove the arm cuff.
Read “Taking a Measurement”.

Wait 2-3 minutes.
Take another measurement.

Quick Start 
Guide

Instruction 
Manual

APPLYING THE ARM CUFF

CAUTION

Read all of the information in the instruction manual and any other
literature in the box before operating the unit.

CAUTION

This device is intended for use in measuring blood pressure and
pulse rate in the adult population. Do not use this device on infants
or persons who cannot express their intentions.

APPLYING THE CUFF ON THE LEFT ARM

1.  Make sure the air plug is securely inserted 

in the main unit.

2.  Remove tight-fitting clothing from your upper 

arm.

3.  Apply the cuff to your left upper arm so 

the arrow is centered on the inside of your arm 
and aligned with your middle finger. 
The air tube runs down the inside of your arm. 
The bottom of the cuff should be 
approximately 1/2 " above your elbow.

4. Wrap the cuff firmly in place around your 

arm using the cloth fastening.

APPLYING THE CUFF ON THE RIGHT ARM

When taking a measurement using the right arm, use this instruction for Step 3.

Apply the cuff to your right upper arm so 
the arrow and the air tube runs along the inside 
of your arm. The bottom of the cuff should be 
approximately 1/2 " above your elbow.

NOTE: 

Be careful not to rest your arm on the air tube. This will restrict the flow of air 
to the cuff.

1/2 inch (1 - 2 cm)

Air Tube

Arrow

Left Arm

Right Arm

Arrow

5325502-1A