GUIDANCE AND MANUFACTURER’S DECLARATION
37
Immunity test
IEC 60601
Test level
Compliance
level
Electromagnetic environment
– guidance
(Table 4)
Guidance and manufacturer’s declaration – electromagnetic immunity
OMRON BPM is intended for use in the electromagnetic environment
specified below. The customer or the user of this OMRON BPM should assure
that it is used in such environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Portable and mobile RF
communications
equipment should be used no
closer to any part of
the OMRON BPM
including cables, than the
recommended separation distance
calculated from the equation
appropriate to the frequency of the
transmitter.
3 V rms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
3 V/m
Not
Applicable.
Recommend
separation distance
Not Applicable.
d
= 1.2 √P
80 MHz to 800 MHz
d
= 2.3 √P
800 MHz to 2.5 GHz