• Before measuring, stretch your back and sit up

straight, breath normally 5-6 times, relax your
shoulders, arms and entire body. 

• Slightly bend and support your elbow (e.g. on a

table). 

• Do not move, talk or touch the device during

measurement. 

• Stress influences your blood pressure, so try to

relax before taking a measurement. 

• To compare results, measurements must always

be taken from the same arm.

• Do not place the cuff over thick clothes and do not

roll up your sleeve if it is too tight.

• Align the cuff with the height of your heart (breast

height). If necessary use a pillow or cushion.

• Coldness raises your blood pressure. It is

recommended to measure at room temperature
(+20°C).

• Relax before measurement. If necessary go to the

bathroom before measurement.

• Do not measure directly after bathing.

• If you wish to interrupt measurement prematurely,

press the (O/I) button. Inflation is stopped and the
air is automatically released from the cuff.

• Blood pressure measurements taken in a moving

vehicle may be influenced by vibration.

• If an error has occurred during measurement 

is

displayed. Please refer to the next chapter.

• The monitor should be switched off after

measurement to avoid running down the batteries
unnecessarily. If you forget to do so, the monitor
switches itself off automatically after 5 minutes.

E

Hints on taking
blood pressure
readings

Failure, causes and rectification

Failure and their possible causes

Incorrect measurement is indicated by 

1. A correct reading could not be obtained

because measurement was disturbed by
movement of the body.

2. The cuff is not fitted correctly.

3. Your clothing has constricted your 

blood flow.

4. There is still air in the cuff when the

monitor is switched off.

The (START

) button was pressed

before the (

) symbol was displayed.

The blood pressure values displayed are
extremely low or high, or they are
implausible.

The display does not light up when the
(START

) button is pressed.

1. The batteries are exhausted.

2. The +/- poles of the batteries have been

reversed.

3. The battery contacts are dirty.

The batteries are weak or exhausted.

The cuff pressure does not rise
although the pump motor can be heard.

problem with mem. function

Er

E

Rectification

1. Repeat the measurement keeping

perfectly still. Do not move your arm and
do not speak.

2. Check that the cuff is correctly fitted.

Then repeat the measurement.

3. Remove the item of clothing which

caused the constriction.

4. The unit may be defective. Please have

it examined and, if necessary, repaired
by OMRON.

Wait for the (

) symbol indicating

readiness before you press the (START

)

button.

Refer to the instructions and then repeat
the measurement.

1. Check the batteries and, if necessary, fit

4 new batteries.

2. Reinsert the batteries with the +/- poles

the right way round.

3. Clean the battery contacts with a dry

cloth.

Fit new batteries.

Check that the air tube is properly
connected to the monitor. Push the tube
connector firmly into the socket.

Contact your Omron retailer or distributor.

Summary of Contents for 705IT

Page 1: ...i per l uso Pagina 86 Gebruiksaanwijzing Pagina 23 Instrucciones de uso P gina 107 Instruction manual Page 2 Mode d emploi Page 65 EU representative EU Vertreter Repr sentant UE Rappresentante UE Repr...

Page 2: ...so helps to support the doctors work The World Hypertension League is a world wide association of experts specialising in hypertension The World Hypertension League recommends regular blood pressure m...

Page 3: ...pter Jack USB Connector Pulse Rate Date Time Display Heart Symbol Memory recall Button Systolic Blood Pressure Set Button Adjust Button Diastolic Blood Pressure Air Jack Slide the battery cover off in...

Page 4: ...nd minutes ADJUST SET DATE TIME SET You can wrap the cuff either on your right or left arm Insert air tube to air jack on the left side of the divice The cuff must be airless Remove tight fitting clot...

Page 5: ...the display Press the START button and remain still As the cuff begins to inflate the monitor automatically determines your ideal inflation level Because this monitor detects the pulse even during in...

Page 6: ...press and hold the START button until the monitor inflates 30 to 40 mmHg higher than your suspected systolic pressure NOTE Do not apply more pressure than necessary The monitor will not inflate above...

Page 7: ...t the mouse pointer to Settings on the menu Click Control Panel 4 Open the Add Remove Programs by double clicking the icon 5 Execute the installation Click the install button Follow the instructions T...

Page 8: ...in your 705IT blood pressure monitor By pressing the CLOSE button you can close the program How to set the time Press the TIME button on your screen The time and date are read from your 705IT blood p...

Page 9: ...ion Failure and their possible causes Incorrect measurement is indicated by 1 A correct reading could not be obtained because measurement was disturbed by movement of the body 2 The cuff is not fitted...

Page 10: ...ovisions of the EC directive 93 42 EEC Medical Device Directive This blood pressure monitor is designed according to the European standard EN1060 Non invasive sphygmomanometers Part 1 General requirem...

Page 11: ...nxiety or the time of day can influence your blood pressure Blood pressure is typically low in the morning and increases from afternoon to evening It is lower in the summer and higher in the winter Bl...

Page 12: ...e blood pressure measured first thing in the morning after getting up before eating and while at rest is known as the fundamental blood pressure In practice it is rather difficult to record the fundam...

Reviews: