035-00708R
R1
3
1. Ensure the AC power cord is disconnected from the unit.
2. Remove the screws from the lamp access side panel using the tool (3mm
allen key) provided and remove the panel from the unit cover.
3. Carefully remove the lamp module from its container, holding only the
ceramic component and glass rim.
Warning: Wear ESD protective equipment when handling the lamp
and avoid touching the circuit board.
4. As illustrated, position the lamp facing towards the front of the unit with the
power leads facing towards you.
Caution: Do not touch the bulb’s glass envelope or the inside
surface of the reflector. Skin oils can cause the lamp module to fail
prematurely.
5. Make sure the middle of the lamp is in position to fit into the spring clamp.
Slide the lamp until it snaps into the spring clamp. The leading edge of the
reflector should fit snugly into the lamp holder recess.
6. Locate the 6-pin Intelli-Lamp
®
sensor connector at the rear of the lamp
module and connect it to its mate located behind the the lamp- housing
wall.
Tip: the Intelli-lamp connector will only attach in the correct orientation. If
you are having difficulty attaching the connector, try rotating it by 180º.
7. Tuck the Intelli-Lamp wires into the grommet located on the divider wall
between the lamp and the connector.
Waring: Failure to do so will result in the door not fully closing and
system will not power on.
8. Locate the Power connector with two leads and connect it to its mate
located on the bottom of the lamp-housing wall.
Tip: The Lamp Power connector will only attach in the correct orientation.
If you are having difficulty inserting the connector, try rotating it by 180º.
9. It is very important to ensure that the lamp anode cooling fin (bar) and
lamp power lead, at the front of the lamp are centered within the lamp
holder cutout. Rotate the lamp as required.
10. Replace the lamp housing side panel and tighten the fastening screws.
Warning: If the lamp housing panel is not secured completely, the
unit will not power on.
Installing the Lamp Module
The OmniCure S2000 Elite curing system is available with the choice of two different lamps. UV1 standard curing
lamp and UV3 Surface curing lamp. Choose the appropriate type of lamp for your curing needs:
Warning: Exposure to Mercury represents a health hazard to humans. When unpacking or installing the
lamp, always wear protective clothing and a facemask.