Omni International Bead Ruptor 4 User Manual Download Page 1

 User Manual

Bead Ruptor

TM

 4

Bead Mill Homogenizer

Summary of Contents for Bead Ruptor 4

Page 1: ...User Manual Bead RuptorTM 4 Bead Mill Homogenizer...

Page 2: ......

Page 3: ...IGENCE AND ALL WARRANTIES OF MERCHANTABILITY OR OTHERWISE EXPRESSED OR IMPLIED IN FACT OR BY LAW AND STATE OUR ENTIRE AND EXCLUSIVE LIABILITY AND BUYERS EXCLUSIVEREMEDYFORANYCLAIMOFDAMAGESINCONNECTION...

Page 4: ...surfaces chauff es NE PAS modifier la fiche ou le cordon fourni Si la fiche ne s adapte pas la prise faites prise appropri e install e par un lectricien qualifi NE PAS utiliser le produit avec la mis...

Page 5: ...box Lift unit up and off of the box bottom DO NOT LIFT THE UNIT BY HOLDING THE LID Remove Tool Kit and Power Cord Place the unit on a stable surface Remove the 2 screws from the bottom of the unit Co...

Page 6: ...eventative Maintenance Normal use 12 month interval Heavy use may require a 6 month interval Electric Supply 4C to 40C 39F to 104F Humidity 5 to 95 RH Internal use only Electric Supply 19 042E 115V 50...

Page 7: ...tative Maintenance Replace Tube holders once a year at a minimum Table of Symbols Symbol Description Symbol Description Caution Refer to the User documentation ISO 7000 0434B On power IEC 60417 5007 H...

Page 8: ...1 Overview 1 4 2 5 3 1 Lid 2 Keypad 3 Lid Latch 4 ON Off Switch 5 Power Cord Port 6 Speed Control 7 Timer Control 8 Run Button 9 Stop Button Speed 5 Time 30 6 8 7 9 The Bead RuptorTM 4 Bead Mill Homo...

Page 9: ...buttons from 1 sec to 5 min 4 Start Press Run button 5 Pause Stop Press Stop Button to Pause Press Stop again to reset timer and cancel run 1 Digital Display 2 Timer 3 Speed Control 4 Timer Control 5...

Page 10: ...Loading Sample Tubes 1 5 mL 2 mL Tubes 7 mL Tube Carriage 7 mL Tubes OMNI Bead tubes should be used exclusively with OMNI Bead RuptorTM Homogenizers...

Page 11: ...the new neck gasket onto the pivot shaft and housing 3 Secure the new neck gasket using the restriction tie with the clamp positioned toward the bottom of the instrument Tighten central screw 4 Use th...

Page 12: ...Unit powers on but motor does not turn Faulty motor Contact manufacturer for assistance Error 4 Close Door Lid is not seated properly Check Lid latch for obstruction Press lid and ensure the handle is...

Page 13: ...will slide back Lift the tube carriage to align the motor mount holes to the holes in the base plate Insert screws with allen key until you feel them engage It may take some finesse to get them align...

Page 14: ...Notes...

Page 15: ...Notes...

Page 16: ...PN 03 225 01 Rev N 935 Cobb Place Blvd Kennesaw GA 30144 800 776 4431 770 421 0058 www omni inc com...

Reviews: