Getting Started with Quick
3 - 9
Power output value
Current (I) or voltage (V) value, depending on the
specified range (see above).
Any value between minimum and maximum as well as
the explicit value 0.0 is possible, even while in the "on"
state.
The respective minimum and maximum values depend
on the range and, for voltage ranges, also on the
frequency. For frequencies smaller than 50Hz, and down
to the minimum frequency of 15Hz, the possible
maximum voltage value decreases linear proportional to
the frequency value (refer to Figure 3, ”Typical maximum
output voltage of 2kV AC output depending on frequency
and power” on page 16-14).
Current ranges do not have such a dependency.
Frequency / Phase angle
Frequency value in Hz or - if "Sync w/ V1AC" is selected
(see below) - phase angle in degrees. For DC ranges this
data entry field is disabled.
The frequency can be set within a range of
15.00 ... 400.00Hz, even while in the "on" state.
The phase angle is displayed in a range of
-180.0 ... +180.0°, and can also be changed at any time.
If this range is exceeded, the phase angle will adapt itself
automatically, that is, change from -180.0 to +180.0°.
Summary of Contents for CPC 100
Page 12: ...CPC 100 V1 41 x...
Page 28: ...CPC 100 V1 41 1 16...
Page 90: ...CPC 100 V1 41 3 14...
Page 194: ...CPC 100 V1 41 6 30...
Page 250: ...CPC 100 V1 41 8 32 Figure 24 Settings of Amplifier test card for this example use case...
Page 258: ...CPC 100 V1 41 9 4...
Page 264: ...CPC 100 V1 41 10 6...
Page 282: ...CPC 100 V1 41 12 10 Figure 9 Saving tests with the CPC Editor...
Page 284: ...CPC 100 V1 41 12 12...
Page 312: ...CPC 100 V1 41 14 12...
Page 316: ...CPC 100 V1 41 15 4...
Page 350: ...CPC 100 V1 41 16 34...
Page 372: ...OMICRON Contact Addresses 22...