Introduction
2 - 9
INPUT
Analog precision measuring inputs
Figure 7:
Functional group INPUT
3.
V2AC
low level AC voltage measuring input
Measuring input for low-level AC voltage of 0...3V in a frequency range of
15...400Hz (for example, for low-power measuring transformers, or to
connect a current clamp).
The software changes automatically between the measuring ranges 0.03V,
0.3V and 3V. For details about the accuracy of the individual ranges, please
refer to ”3V V2 AC Input” on page 16-25.
4. VDC voltage measuring input
10V DC
Measuring input for DC voltage of 0...10V. The software changes
automatically between the measuring ranges 0.01V / 0.1V / 1V / 10V.
Note:
The
V1 AC
and
V2 AC
inputs are galvanically connected. All other
inputs and outputs are separated.
1.
IAC/DC
current measuring input 10A AC and DC
Measuring input for AC and DC current of 0...10A,
depending on the selected test card. AC current is
measured in a frequency range of 15...400Hz.
For both AC and DC current measurement, the
software changes automatically between the two
measuring ranges 1A and 10A. In both ranges the
measuring error is <±0.2%.
A
10A FF
(10 Amps very quick-acting wire fuse
6.3x32mm) protects the 0...10A AC/DC current
measuring input.
2.
V1AC
voltage measuring input 300V AC
Measuring input for AC voltage of 0...300V. AC voltage
is measured in a frequency range of 15...400Hz.
During measurement, the software changes
automatically between the measuring ranges 0.3V, 3V,
30V and 300V.
For details about the accuracy of the individual ranges,
please refer to ”300V V1 AC Input” on page 16-23.
Summary of Contents for CPC 100
Page 12: ...CPC 100 V1 41 x...
Page 28: ...CPC 100 V1 41 1 16...
Page 90: ...CPC 100 V1 41 3 14...
Page 194: ...CPC 100 V1 41 6 30...
Page 250: ...CPC 100 V1 41 8 32 Figure 24 Settings of Amplifier test card for this example use case...
Page 258: ...CPC 100 V1 41 9 4...
Page 264: ...CPC 100 V1 41 10 6...
Page 282: ...CPC 100 V1 41 12 10 Figure 9 Saving tests with the CPC Editor...
Page 284: ...CPC 100 V1 41 12 12...
Page 312: ...CPC 100 V1 41 14 12...
Page 316: ...CPC 100 V1 41 15 4...
Page 350: ...CPC 100 V1 41 16 34...
Page 372: ...OMICRON Contact Addresses 22...