Resistance
7 - 9
Winding Resistance
Never open the measuring circuit while current flows. Dangerous voltage may
occur!
Figure 6:
Setup for a winding
resistance test
Use the
RWinding
test card to measure the resistance of a current
transformer’s secondary winding.
To do so, open the circuit as shown in Figure 6, loop the I DC current signal from
CPC 100
6A DC
output via
IAC/DC
or directly from
400A DC
output, and inject
it into the transformer’s secondary winding.
Note:
Some test cards are available in more than one test mode. For
example, the test card
RWinding
can be selected in CT, Resistance
and Transformer.
This is solely related to usability. With regard to functionality, the test
card
RWinding
is identical in all test modes.