6.6 Pulse Out 2
Interface
Figure 6-2:
Pulse out 2
interface
Figure 6-3:
Block diagram of the
Pulse out 2
interface
Table 6-6:
Technical specifications
Pulse out 2
interface
The synchronization pulse is only put out if the
CMGPS
is able to meet the
accuracy configured by the software.
At this interface the synchronization pulse of the
CMGPS
can be picked up potential free via an
internal opto-electronic coupler (open collector). At
present, the
Test Universe
software supports the
configuration of a pulse output only. Therefore,
Pulse out 2
is synchronous to
Pulse out 1
.
+ U
B
(max. 30 V)
30 V
Pull-up resistor
R
p
(optional)
Pulse out 2
Bin In +
GND
red +
black -
CMGPS
CMC test set
Interface Pulse out 2
Type
Output by optoelectronic coupler (open collector)
Accuracy
Absolute error < ±5 µs
1
1.
Depending on accuracy set in the
Test Universe
software
Pulse length
200 ms
Pulse rate
Configurable in steps of 1 s (1 s … 65535 s)
Polarity of synch. signal
Configurable to falling or rising edge
First synchronization point
of time
Programmable
Specifications of the open
collector output transistor:
•
I
Cmax
•
U
CEmax
5 mA
30 V (red , black socket –)
Summary of Contents for CMGPS
Page 1: ...27 CMGPS Synchronization Unit Reference Manual...
Page 8: ...GPS Synchronization Unit CMGPS 8...
Page 10: ...GPS Synchronization Unit CMGPS 10...
Page 24: ...GPS Synchronization Unit CMGPS 24...
Page 28: ...GPS Synchronization Unit CMGPS 28...
Page 34: ...GPS Syncronization Unit CMGPS 34...