![Omicron CMC 356 User Manual Download Page 112](http://html2.mh-extra.com/html/omicron/cmc-356/cmc-356_user-manual_3980661112.webp)
CMC 356 Reference Manual
112
9.5 Current Clamp C-PROBE1
Using the current clamp
C-PROBE1
and the option ELT-1, direct and
alternating currents can be measured via the analog measurement inputs of
the BINARY / ANALOG INPUT section (refer to section 6.10, "Option ELT-1"
on page 67).
C-PROBE1
is an active, DC-capable current probe and has two switchable
measurement ranges.
Figure 9-9:
Current clamp
C-PROBE 1
For detailed information about the
C-PROBE1
current clamp and the option
ELT-1, please refer to the respective reference manuals, the product
catalog, or visit the OMICRON Web site
www.omicron.at
.
Table 9-4:
Basic technical data of
C-PROBE 1
For ordering information about the
CPROBE-1
, please refer to table 9-5,
"Order numbers overview" on page 128.
DC / AC
"+" = red
"-" = black
Max. voltage of the leads 600 V
rms
to GND
Switch position
100 mV/A
10 mV/A
Measurement ranges
0…10 A AC/DC
0…80 A AC/DC
Frequency bandwidth
0 (DC)…10 kHz