Technical Data
67
6.10 Option
ELT-1
Figure 6-18:
Binary/analog inputs and
inputs for analog DC
measurement
The ELT-1 option enables the
CMC 356
to measure analog signals:
• Analog DC inputs (+/-10V and /-1mA or +/-20mA) for basic
transducer testing with the test module
QuickCMC
.
• Basic voltage and current measurements with up to three of the 10
analog measurement inputs (restricted
EnerLyzer
mode).
In addition, the
Test Universe
module
EnerLyzer
provides the following
functionality:
• Simultaneous measurement of up to 10 voltages and/or currents.
• Evaluation of DC components (DC voltages or DC currents).
• Real-time indication of effective values (true RMS) for all measurement
signals.
• Peak values indication (U
peak
, I
peak
,...).
• Phase angles with reference to a given input signal.
• Real-time calculation of apparent, reactive, and active power (in any
configuration).
• Frequency and spectrum indication (harmonic diagrams) of periodic
signals.
• Capturing of transient input signals with different sampling rates.
• Different triggering options for transient signal capturing (basic triggers
and power quality triggers).
• Trend Recording: Measurement of RMS current, RMS voltage,
frequency, phase, active, apparent and reactive power and power factor
over long periods of time (up to 4 million samples possible).
Binary/analog inputs
Analog DC Inputs
Summary of Contents for CMC 356
Page 1: ...CMC 356 User Manual...
Page 38: ...CMC 356 Reference Manual 38...
Page 100: ...CMC 356 Reference Manual 100...
Page 104: ...CMC 356 Reference Manual 104...
Page 132: ...CMC 356 Reference Manual 132...
Page 136: ...The OMICRON Bootloader Software Copyright Notices 136...