Troubleshooting
103
8.3 Overheating
If a thermal shutdown occurs because of loading the voltage or current
outputs a long time by high burden, the
Test Universe
displays the following
messages respectively in the Status History window:
• “Voltage overtemperature:” followed by a list of the affected
outputs
“CMC switched off.”
“Test stopped with error.”
• “Current overtemperature:” followed by a list of the affected
outputs
“CMC switched off.”
“Test stopped with error.”
The thermal shutdown can be avoided by reducing the compliance voltage
of the current amplifiers, i.e., to optimize the output power limit of the current
outputs set the compliance voltage of the internal current amplifiers.
To do so, go to the
Compliance Voltage
group box of the
Output
Configuration Details
dialog box in the
Test Universe
Hardware
Configuration
.
By reducing the power supply voltage, the ON-time can be prolonged
considerably for low-ohmic burdens, because this causes the internal
amplifier to consume less power. Hence, the internal heat dissipation can be
reduced, especially when testing with low burden test objects. This then
considerably extends the time until the device switches OFF due to thermal
overload.
For more detailed information refer to the
Test Universe
Help. Select the
Hardware Configuration
Help and navigate to the topic
Setting the
Current Output Power Limit of CMC Test Sets
.