Device description
17
Figure 6-2
SPECTANO 100
rear view
DANGER
Death or severe injury caused by high voltage or current
►
Do not handle
SPECTANO 100
without grounding it.
►
Before handling
SPECTANO 100
in any way, connect its equipotential
ground with a solid connection of at least 6 mm
2
cross-section and not
longer than 20 meters to a ground connection in the laboratory.
Death or severe injury caused by high voltage or current
►
Do not use power outlets without protective ground (PE).
►
Only use the delivered grounded AC power supply
DRA SUPPLY
to power
SPECTANO 100
.
►
Connect the
SPECTANO 100
power supply only to a power outlet that is
equipped with protective ground (PE).
►
For safe operation always make sure that protective ground and
equipotential ground are connected properly.
DANGER
USB connector
For connecting
SPECTANO 100
to a computer
with a USB 2.0 A/A cable
Equipotential ground
For connecting an external ground
I/0 (on/off) power switch
5 V DC output
Optional: for connecting a USB isolator
DC power input
Connect only the especially designed and
optimized delivered power supply
DRA SUPPLY
for testing with
SPECTANO 100
.