Default Standard
Select the default I
sn
, standard, class and CT type (P/M)
setting to be used for a new CT test using the soft keys.
"Isn": Select the default secondary current (
1A
or
5A
, or
?
if you want to use the guesser function by default).
"Standard": Select
IEC 61869-2
,
IEC 60044-1
,
IEC 60044-6
or
IEEE C57.13
.
"P/M": Select
Prot. CT
or
Meter. CT
(protection or
metering CT), or
?
if you want to use the guesser
function by default.
"Class": The classes available for selection depend on
the selected standard and CT type (P/M). Select a class
or
?
if you want to use the guesser function by default.
"Ktd" (IEC 60044-6 only):
If IEC 60044-6 is selected for the standard, it is
furthermore possible to select the calculation method for
the K
td
(transient dimensioning factor):
acc. to IEC 60044-6
: K
td
is calculated exactly according
to the standard.
acc. to OMICRON
: K
td
calculation considers the CT
remanence.
acc. to GB 16847
: K
td
is calculated acc. to the Chinese
standard GB16847.
"TPX/Y/Z Spec. Mtd." (IEC 61869-2 protection CTs only):
For IEC 61869-2 protection CTs of the classes TPX, TPY
and TPY it is possible to select the default method for
specifying the protection requirements used for the
accuracy class assessment (refer to page 93 for more
information).
by Duty
: Protection requirements specification by duty
cycle and the necessary time constants (standard
specification according to IEC 61869-2; the
CT Analyzer
then calculates the K
td
automatically from the entered
values).
by Ktd
: Protection requirements specification by the
rated transient dimensioning factor K
td
stated on the
CT’s name plate (alternative specification according to
IEC 61869-2).
Default Nominal
Frequency
Nominal frequency to be used for a new CT test:
16.7Hz
,
50Hz
,
60Hz
or
400Hz
.
Summary of Contents for CT Analyzer
Page 1: ...CT Analyzer CT SB2 User Manual...
Page 3: ...CT Analyzer User Manual...
Page 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Page 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Page 50: ...CT Analyzer User Manual 48...
Page 60: ...CT Analyzer User Manual 58...
Page 80: ...CT Analyzer User Manual 78...
Page 162: ...CT Analyzer User Manual 160...
Page 172: ...CT Analyzer User Manual 170...
Page 188: ...CT Analyzer User Manual 186...
Page 194: ...CT Analyzer User Manual 192...
Page 195: ...CT SB2 User Manual...
Page 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Page 204: ...CT SB2 User Manual 10...
Page 206: ...CT SB2 User Manual 12...
Page 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Page 212: ...CT SB2 User Manual 18...
Page 226: ...CT SB2 User Manual 32...
Page 242: ...CT SB2 User Manual 48...
Page 248: ...CT SB2 User Manual 54...
Page 250: ...CT SB2 User Manual 56...
Page 252: ...CT SB2 User Manual 58...