29
Setup and Connection
Figure 2-6
Using a copper foil formed to a ring as primary wire
Note:
Exact measurement results are only possible if the primary wire is
positioned exactly in the center of the core.
The
CT Analyzer
does not consider leakage inductances. The leakage
inductances are neglected instead. This way, the
CT Analyzer
is able to reach a
measurement error of approx. 0.1% for class PR and class TPY CTs and
approx. 0.8% for class TPZ CTs.
2.5.2
Excitation Curve Measurement for an Unwound Iron Core
Using the
CT Analyzer
it is possible to measure the magnetic properties of an
empty, unwound iron core. For this purpose, it is necessary to apply an "auxiliary
winding" of at least 20 turns to the core.
For this purpose, OMICRON offers a special cable with 23 turns (VEHK0658)
and a special Microsoft Excel template for the required calculations.
Proceed as follows to perform the measurement (refer to Figure 2-7 and Figure
2-8). Observe the general safety rules in section 2.1 on page 21.
1. Apply the "auxiliary winding" cable to the unwound core.
2. Connect the "Output" sockets and input "Sec" of the
CT Analyzer
to the cable
as shown in Figure 2-7. For cores that require high currents to reach the knee
point, several cables can be cascaded to increase the number of turns, see
Figure 2-8.
3. Start the
CTA Remote Excel File Loader
with the special Excel template
mentioned above.
4. Enter the iron parameters to the template.
Copper foil
Summary of Contents for CT Analyzer
Page 1: ...CT Analyzer CT SB2 User Manual...
Page 3: ...CT Analyzer User Manual...
Page 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Page 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Page 50: ...CT Analyzer User Manual 48...
Page 60: ...CT Analyzer User Manual 58...
Page 80: ...CT Analyzer User Manual 78...
Page 162: ...CT Analyzer User Manual 160...
Page 172: ...CT Analyzer User Manual 170...
Page 188: ...CT Analyzer User Manual 186...
Page 194: ...CT Analyzer User Manual 192...
Page 195: ...CT SB2 User Manual...
Page 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Page 204: ...CT SB2 User Manual 10...
Page 206: ...CT SB2 User Manual 12...
Page 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Page 212: ...CT SB2 User Manual 18...
Page 226: ...CT SB2 User Manual 32...
Page 242: ...CT SB2 User Manual 48...
Page 248: ...CT SB2 User Manual 54...
Page 250: ...CT SB2 User Manual 56...
Page 252: ...CT SB2 User Manual 58...