21
Setup and Connection
2
Setup and Connection
2.1
General Safety Rules for Connecting and Operating
the CT Analyzer
Observe the following general safety rules when connecting and operating the
CT Analyzer
. The safety rules given here are supplemented by further notes and
warnings applicable for specific actions only. Such specific notes and warnings
are given where necessary in this user manual.
• Before putting the
CT Analyzer
into operation, check the test set for visible
damages.
• When taking the
CT Analyzer
into operation, make sure that the air slots, the
power switch and the power supply plug at the test set are not obstructed.
• Only use wires with 4mm safety "banana" connectors and plastic housing for
connection to the front panel input/output sockets.
• During the test always connect one terminal of the transformer’s primary side
to protective earth.
Warning:
Make sure that the terminals of the test object to be connected to the
CT Analyzer
do not carry any voltage potential. During a test, the
CT Analyzer
is the only permitted power source for the test object.
Warning:
Do not touch the equipment under test or the measurement leads
while the red LED on the
CT Analyzer
is flashing. Never connect or disconnect
measurement leads while the red LED on the
CT Analyzer
is flashing.
As long as the red LED is flashing, the output is active and lethal voltages can
occur due to the high energy stored in external inductors.
Warning:
Always make sure that the
CT Analyzer
output is connected to the
correct side of the current transformer according to the wiring instructions given
in sections 2.4 to 2.5 below. Accidentally mixing up the windings can cause life-
threatening voltages within the transformer and/or destroy the connected CT
or the
CT Analyzer
!
Summary of Contents for CT Analyzer
Page 1: ...CT Analyzer CT SB2 User Manual...
Page 3: ...CT Analyzer User Manual...
Page 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Page 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Page 50: ...CT Analyzer User Manual 48...
Page 60: ...CT Analyzer User Manual 58...
Page 80: ...CT Analyzer User Manual 78...
Page 162: ...CT Analyzer User Manual 160...
Page 172: ...CT Analyzer User Manual 170...
Page 188: ...CT Analyzer User Manual 186...
Page 194: ...CT Analyzer User Manual 192...
Page 195: ...CT SB2 User Manual...
Page 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Page 204: ...CT SB2 User Manual 10...
Page 206: ...CT SB2 User Manual 12...
Page 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Page 212: ...CT SB2 User Manual 18...
Page 226: ...CT SB2 User Manual 32...
Page 242: ...CT SB2 User Manual 48...
Page 248: ...CT SB2 User Manual 54...
Page 250: ...CT SB2 User Manual 56...
Page 252: ...CT SB2 User Manual 58...