![Omicron Lab CT Analyzer User Manual Download Page 26](http://html2.mh-extra.com/html/omicron-lab/ct-analyzer/ct-analyzer_user-manual_3973450026.webp)
CT Analyzer User Manual
24
2.4
Connection for Usual Applications
2.4.1
Wiring for a CT Test
This section shows the basic wiring of the
CT Analyzer
for a CT test.
Observe the general safety rules in section 2.1 on page 21 and the hints and
instructions for improving the quality of the measurement results given in section
2.7 on page 42.
For a CT test, connect the
CT Analyzer
as shown in Figure 2-2:
1. Make sure that the primary side of the CT is connected to PE on one side and
open on the other side.
2. Disconnect the hot side of all secondary windings of the CT in order to
remove any load from the CT. Every kind of load remaining on the secondary
side of the CT during measurement leads to incorrect measurement results
or error messages.
3. Connect the black socket of
CT Analyzer
input "Prim" to the grounded side
of the CT’s primary circuit and the red socket of this input to the open
(ungrounded) side.
Warning:
Feeding test voltage to a tap of a
multi-ratio CT
can cause life-
threatening voltages on other taps with higher ratios. Do not touch these taps!
Caution:
The grounded terminal on the secondary side of the CT always has
to be connected to the black sockets of input "Sec" and the "Output" of the
CT Analyzer
. Connecting the red sockets of the
CT Analyzer
to PE can result
in incorrect measurement and/or cause an automatic abortion of the
measurement with an error message.
Caution:
If it is necessary to use clamps for the connection of the
measurement leads to the secondary side of the test object, always use the
4-wire connection technique as described in section 2.7.1 on page 42 in order
to avoid measurement errors.
Caution:
It is absolutely important to avoid coupling of interferences into the
primary circuit during measurement. Therefore, connect the side of the primary
circuit that is able to receive more interferences to PE (e.g. the side with the
longer line length). The ungrounded side should be the side that receives less
interferences (refer to Figure 2-2).
Summary of Contents for CT Analyzer
Page 1: ...CT Analyzer CT SB2 User Manual...
Page 3: ...CT Analyzer User Manual...
Page 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Page 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Page 50: ...CT Analyzer User Manual 48...
Page 60: ...CT Analyzer User Manual 58...
Page 80: ...CT Analyzer User Manual 78...
Page 162: ...CT Analyzer User Manual 160...
Page 172: ...CT Analyzer User Manual 170...
Page 188: ...CT Analyzer User Manual 186...
Page 194: ...CT Analyzer User Manual 192...
Page 195: ...CT SB2 User Manual...
Page 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Page 204: ...CT SB2 User Manual 10...
Page 206: ...CT SB2 User Manual 12...
Page 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Page 212: ...CT SB2 User Manual 18...
Page 226: ...CT SB2 User Manual 32...
Page 242: ...CT SB2 User Manual 48...
Page 248: ...CT SB2 User Manual 54...
Page 250: ...CT SB2 User Manual 56...
Page 252: ...CT SB2 User Manual 58...