![Omicron Lab CT Analyzer User Manual Download Page 227](http://html2.mh-extra.com/html/omicron-lab/ct-analyzer/ct-analyzer_user-manual_3973450227.webp)
33
Default Settings on the CT Analyzer for Multi-Ratio CT Testing
5
Default Settings on the CT Analyzer for
Multi-Ratio CT Testing
The
CT Analyzer
allows customization of the default settings for multi-ratio CT
testing using the
CT SB2
. These settings are defined using the
Select Startup Mode
option in the
Setting Menu
of the
CT Analyzer
.
Open the
Set Startup Mode
page and press the
Multi-Ratio
soft key. The
Set Startup Mode
page then looks as shown in Figure 5-1.
Figure 5-1
Set Startup Mode
page with
Multi-Ratio
startup mode selected
Modify the default settings according to your needs and press the
Back
soft key
to apply your default settings.
Mode at Startup
According to the mode selected, the
CT Analyzer
automatically initializes a single tap CT test, a multi-ratio
CT test or a QuickTest measurement after startup.
Select the mode using the soft keys
Single Tap
,
Multi-Ratio
or
Quick
.
If the
Multi-Ratio
mode is selected, the CT test initialized
after switching on the
CT Analyzer
contains the
additional test cards
MR-Config.
and
MR-Results
for
multi-ratio CT testing.
Number of Taps
The number of taps is the overall number of tap
connections available on the multi-ratio CT.
Select the default number of taps using the soft keys
No. of Taps 2
to
No. of Taps 6
.
The default number of taps defined here is used each
time a new multi-ratio CT test is initialized by selecting
"New MR-Test" from the main menu of the
CT Analyzer
.
Press the
Main
soft key in
any test card on the
CT Analyzer
Main Menu:
- Settings
Select
soft key
Setting Menu:
- Select Startup Mode
Select
soft key
->
Set Startup Mode
page
How to get there:
Summary of Contents for CT Analyzer
Page 1: ...CT Analyzer CT SB2 User Manual...
Page 3: ...CT Analyzer User Manual...
Page 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Page 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Page 50: ...CT Analyzer User Manual 48...
Page 60: ...CT Analyzer User Manual 58...
Page 80: ...CT Analyzer User Manual 78...
Page 162: ...CT Analyzer User Manual 160...
Page 172: ...CT Analyzer User Manual 170...
Page 188: ...CT Analyzer User Manual 186...
Page 194: ...CT Analyzer User Manual 192...
Page 195: ...CT SB2 User Manual...
Page 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Page 204: ...CT SB2 User Manual 10...
Page 206: ...CT SB2 User Manual 12...
Page 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Page 212: ...CT SB2 User Manual 18...
Page 226: ...CT SB2 User Manual 32...
Page 242: ...CT SB2 User Manual 48...
Page 248: ...CT SB2 User Manual 54...
Page 250: ...CT SB2 User Manual 56...
Page 252: ...CT SB2 User Manual 58...