the electrode to see if the electrolyte in the electrode is run out. If yes, please fill the electrolyte and make 

the new calibration. 

 

 

Electrode Cleaning Techniques 

When your pH or ORP electrode gets slow, fouled and coated with contaminants, there are several things 

that you can do to restore performance. The reason for this is that the sensing glass membrane becomes 

coated. These coatings range from organics such as oil, grease, proteins to metal ions or other 

contaminants. Removing them usually requires soaking or washing in a solution which will remove the 

coating; similar to cleaning your eyeglasses. 

 

a) Cleaning metal ions, some proteins and organics 

    Soaking the electrode in 0.01N. HCl for 10-20 minutes. Follow this by soaking in pH electrode storage solution 

    for 1-2 hours. Rinse the electrode in deionized water, as usual, and recalibrate. 

b) Cleaning grease and oils 

Washing with acetone or methanol and then warm tap water and dishwashing liquid. Use care if cleaning 

mechanically as the pH or ORP glass is fragile. Electrodes washed in this manner also need to be   

conditioned by storing in pH/ORP electrode storage solution and recalibrated. 

c) Salt Build-Up 

Salt crystals can accumulate when water in a salt solution evaporates during contact with air. These salt   

crystals can be removed by rinsing the area with warm water. 

 

TROUBLE SHOOTING 

Q1:    pH display erratic readings or calibration errors?? 

A1:   

Air Bubble in Glass Bulb 

During shipment, horizontal storage, and from general use, sometimes   

allowing air bubbles into the glass Bulb. Before each use, it is recommended 

to inspect that the glass bulb is no visible air bubbles.If there are, shake the   

electrode in a downward motion to force air bubbles out of the glass bulb. 

Dirty Glass Bulb 

Some contaminants remain on the glass surface. Refer to page 9 for specific guidelines for the 

most suitable cleaning solution for the specific contamination you have. 

Un-plugged the electrode 

The readings keep changing on the screen are not referable. 

Q2:    Wrong temperature?? 

A2:   

Refer to page 4 (

pH TEMPERATURE PROBE TYPE

), you must select the correct temperature sensor 

type or adjust temperature manually (Long press 

UNIT

 button then press 

UNIT

 

to select “

not

”). 

Q3:    Appear “Error codes”?? 
A3:   

Make sure if Hold is appeared on LCD. 

ERROR CODES 

Code 

Description 

- - - 

The electrode was not in a solution. Dashes in the temp. mode indicates that probe without 

ATC or manual temp. setting. The electrode of cond. measured in pure water or ultra-pure water. 

Select the wrong temp. electrode type (refer to page 4). 

OL2 

Measurement is out of range of the display. 

 

Summary of Contents for PHH444

Page 1: ...e mail info omega com For latest product manuals omega com en us pdf manuals Shop online at omega com User sGuide PHH444 Handheld pH mV Meter and pH Electrode Kit TM 2...

Page 2: ...pecifications without notice Servicing North America U S A Omega Engineering Inc Headquarters Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Canada only Engineering Servi...

Page 3: ...brated before shipment Warranty period of meter is two years Electrode is one year SUPPLIED Meter Electrode x 1 pc pH Battery AAA x 3 pcs pH buffer solution x 3 pcs pH 4 pH7 pH10 Soaking Solution x 1...

Page 4: ...rt the braid into the center of the connector Continuing pushing the braid into the connector until it will not go in any farther Do this gently and slowly do not bend the braid c Turn the BNC connect...

Page 5: ...measurement Meter default auto power off 6 FACTORY RESET a pH electrode Whenever replacing a new pH electrode Long pressing MODE CAL to enter pH calibration mode then strongly recommend doing factory...

Page 6: ...ibrating the pH in the incorrect buffer solution Er 3 will appear on the bottom of screen 4 User can do 1 2 or 3 points calibration Rinse the probe with de ionized water before doing each pH buffer ca...

Page 7: ...NIT button to select not type HOLD to increase ADJ to decrease SET to left digit CAL to right digit Long pressing UNIT again to save setting until screen shows SA and back to normal measurement mode 4...

Page 8: ...umulate when water in a salt solution evaporates during contact with air These salt crystals can be removed by rinsing the area with warm water TROUBLE SHOOTING Q1 pH display erratic readings or calib...