26
5.3
P03:
OUTPUT ELECTRIC CIRCUIT
(
SP1/SP2
)
SUB-FUNCTION
done
done
HOLD
SET
HOLD
SET
HOLD
SET
HOLD
SET
HOLD
SET
HOLD
SET
HOLD
SET
HOLD
SET
HOLD
ESC
ESC
ENT
ENT
ENT
ENT
ENT
ENT
ENT
7.00
SPI
ATC
O
C
SP2
I2.00 25.0
MEA
PH
7.00
SPI
ATC
O
C
SP2
I2.00 25.0
MEA
PH
1.
Press
to confirm in P03 sub-function. Entering set up of the sub-function.
2.
The lower display shows CTYP. The upper display shows 4 (it means electric circuit output
from 4 m A to 20 m A). User can press
or
select 0 (it means electric circuit output
from 0 to 20 m A). Press
to confirm and enter into set-up of transmitting range.
3.
The lower display shows CURL while the upper display shows 0.00, which means that
0.00mA transmitting range indicates 0.00pH. User can press
or
to adjust the actual
value. Press
to confirm. The lower display shows CURH while the upper display
shows14.00, which means 20.00mA transmitting range indicates 14.00pH. User can press
or
key to adjust the actual value. Press
to confirm and return to P03 sub-
function firm. Use
or
to select other sub-functions and to set up.
For example, set CTYP to 0 or 4, set CURL to 2.00 and set CURH to 10.00, which means 0 or
4 to 20mA output mode is selected as circuit output. 0/4.00mAindicates 2.00pH, 20.00mA
indicates 10.00pH
°
Note
: Press
to quit set-up mode at anytime. Instrument will return to measurement
mode automatically.
Note
:
This parameter allows you to set range of circuit output, but
transmitting high point and low point can’t be overlapped.