Ref.No.
:
062017
STORAGE & CLEANING
It should be stored at room temperature. The sensor lens is the most delicate part of the thermometer. The lens should be kept clean at all times,
care should be taken when cleaning the lens using only a soft cloth or cotton swab with water or medical alcohol, allowing the lens to fully dry
before using the thermometer. Do not submerge any part of the thermometer.
LCD ERROR MESSAGES
The thermometer incorporates visual diagnostic messages as follows:
‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the settings of HAL and LAL.
「
」
「
」
‘Er2’ is displayed when the thermometer is exposed to rapid changes in the ambient temperature. ‘Er3’
is displayed when the ambient temperature exceeds 0
o
C (32°F) or +50
o
C (122°F). The thermometer
should be allowed plenty of time (minimum 30 minutes) to stabilize to the working/room temperature.
「
」
Error 5~9
, for all other error messages it is necessary to reset the thermometer. To reset it, turn the
instrument off, remove the battery and wait for a minimum of one minute, reinsert the battery and turn
on. If the error message remains please contact the Service Department for further assistance.
「
」
「
」
‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the measurement range.
BATTERIES
The thermometer incorporates visual low battery indication as follows:
‘Battery OK’: measurements
are possible
‘Battery Low’: battery needs to be
replaced, measurements are still possible
‘Battery Exhausted’: measurements are not
possible
When the ‘Low Battery’ icon indicates the battery is low, the battery should be replaced immediately with AA, 1.5V batteries. Please note: It is
important to turn the instrument off before replacing the battery otherwise the thermometer may malfunction.
Dispose of used battery promptly and keep away from children.
SPECIFICATION
Item
Non-contact Infrared Scan function
Thermocouple Probe Scan function
(K type; probe not included.)
Measurement Range
-35 to +1600
o
C (-31 to +2912
o
F)
-64 to +1400°C (-83.2 to +2552°F)
Operating Range
0 to +50°C (32 to +122°F)
Accuracy(Tamb=23±3°C) Tobj=-35~0: ±(2°C+0.05/degC)°C,
Tobj=0~1600: ±2% of reading or 2°C (4°F) whichever is greater
±1% of reading or 1°C (1.8°F) whichever is
greater (Test under Tamb=23±6
°
C)
Emissivity Range
0.95 default – adjustable 0.1 to 1 step .01
Resolution
0.1°C/0.1°F at -83.2 to 999.9(°C/°F), otherwise 1°C/1°F
Spectral Range
8~14
μ
m
Response Time (90%)
1sec
Distance:Spot
60:1 (90% energy covered)
Internal Memory
Detailed 24 Memories with Temperature & Emissivity
External Memory Type
Support 8G micro SD card (Recommend to use the attached SD card).
If SD card data logger is failed, please reset the switch
○
11
USB Function
Please download the program from CD. (CD & USB cable are enclosed with the package)
Output Interface
SPI, USB, SD card
Battery Life
Typical 14, min 10 hours continuous use (Alkaline, with laser and backlight)
Dimensions
202.99*176.07*58.31 mm (7.99×6.93×2.30 inch)
Weight
401 grams (14.14 oz) including batteries (AA*2pcs)
Note: Under the electromagnetic field of 3V/m from180 to600 MHz, the maximum error is 10
o
C (18
o
F).
Caution: The measure range is for thermometer only. User should choose proper probe types for different kinds of application.
Please make sure the target to be measured will not exceed the temperature range of the probe to avoid permanent damage of
the thermocouple probe.
Caution: To avoid electric shock and thermometer damage, do not measure live circuit where voltage exceeding 24V AC RMS or
60V DC with the thermocouple probe.
EMC/RFI:
Readings may be affected if the unit is operated within radio frequency electromagnetic field strength of
approximately 3 volts per meter, but the performance of the instrument will not be permanently affected.