background image

!

FIGURE 1. OMNI-AMP IV Block Diagram

Features

n

100% Encapsulated

n

2 µV/°C Zero Stability

n

0.025°F/°F Reference Junction Stability

n

Up or Down Scale TC Burn-out

n

115 VAC or 230 VAC Power

n

Easy Mounting

n

±10 VDC Output

n

Input Protection to 115 VAC or DC

n

Scale Factor of 10 mV/°F

n

Over 120 dB CMRR at 60 Hz

Description

The OMNI-AMP IV is a high performance rugged epoxy
encapsulated thermocouple amplifier with built-in power
supply. It can be used with Type E, J, T, or K thermocouples
by changing only the span adjustment. Types R, S, and B can
be accommodated by ordering the 1 mV/°F scale factor
option. Only the highest quality industrial components are
used for reliable and long operation. Connections are made
with an ordinary screwdriver to rugged terminal strips. The
amplifier can be mounted using the set of bolts supplied with
the unit.

A reference junction temperature output is also supplied.
Figure 1 shows a block diagram of the amplifier.

Amplifier

The amplifier is a multiple IC instrumentation amplifier with
excellent zero and gain stability. Gain can be adjusted from 40
to 1500 so the amplifier can be used with a variety of
thermocouple signals. An internal path is provided for amplifier
input bias current. The amplifier has excellent common mode
rejection allowing the thermocouple input leads to be several
hundred feet long without degraded performance due to 60
Hz common mode pick-up. The amplifier is also protected
against an accidental over-voltage applied to the input. Please
note that the amplifier has cold junction compensation but  not
linearization. The maximum output is ±10 Volts, providing up
to 1000°F output at 10 mV/°F or 10,000°F at 1 mV/°F.

OMNI-AMP IV

Summary of Contents for OMNI-AMP IV Series

Page 1: ...An OMEGA Technologies Company 7IAH I KE A www omega com e mail info omega com OMNI AMP IV Series Thermocouple Amplifier...

Page 2: ...4 06342 e mail france omega com Germany Austria Germany Austria Germany Austria Germany Austria Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany Tel 49 07056 3017 FAX 49 07056 8540 Toll...

Page 3: ...er to rugged terminal strips The amplifier can be mounted using the set of bolts supplied with the unit A reference junction temperature output is also supplied Figure 1 shows a block diagram of the a...

Page 4: ...input thermocouple and replace with a tin plated copper wire short 2 Jumper the ambient temperature output terminal to common 3 Observe the output voltage with a 4 digit DVM n i a G n a p S t u p t u...

Page 5: ...he gain near the temperature of most interest to obtain the best accuracy around that temperature First adjust the zero and leave the ambient temperature output terminal jumper to common Look in a tab...

Page 6: ...l a i r e t a M c i t e n g a M s d l e i F c i t e n g a M r o c i r t c e l E c i t a t S y l l a n o i s a c c O e g a t l o V C D h g i H s e c r u o S t n e r r u C r o e r a s d l e i f s s e l...

Page 7: ...se shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by OME...

Page 8: ...addlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conductivity Equipme...

Reviews: