![background image](http://html1.mh-extra.com/html/omega/omb-daqboard-3000-series/omb-daqboard-3000-series_user-manual_3930526062.webp)
An Example of Timing Mode
The following example represents one channel in timing mode. The time desired is between the
rising
edge
on the
input channel
and the
falling edge
on the
mapped channel
. Zeroes are returned, in the scan,
until one complete time measurement has been taken. At that point, the value (time in ticks) is latched and
logged by the /3000 Series board until the next time measurement has been completed. Rising edges on the
input channel will clear the counter and falling edges on the mapped channel will latch the output of the
counter at that time. If the scan period is much slower than the rate of time-frames coming [available on the
two channels] then the data will miss some time-frames. The scan period can be decreased to capture more
time-frames.
The data returned is interpreted as time measured in ticks. This data represents the number of ticksize
intervals counted during the timing measurement. There are four timebase settings: 20.833 ns, 208.33 ns,
2.083 µs, and 20.83 µs. These are often referred to as tick-sizes. The 3000 Series board uses a 48 MHz,
30 ppm oscillator as a timing source.
If the input signal has a poor slew rate the timing mode will provide variant results,
dependant upon the input switching threshold.
Example of One Channel in Timing Mode
TIMING: OPT2:
This determines whether the time is to be measured with a
16-bit
(Counter Low), or
32-bit
(Counter High) counter. Since time measurements always have the “stop at the top” option enabled,
this option dictates whether the measurement has a range of 0 to 65535 ticks or 0 to 4,294,967,295 ticks.
5-14 Counter Input Modes
918494
DaqBoard/3000 Series User’s Manual
Summary of Contents for OMB-DAQBOARD-3000 Series
Page 7: ...This product requires one of the following Operating Systems Windows 2000 Windows XP ...
Page 14: ...IG 8 DaqBoard Installation Guide 918494 1033 0940 rev 8 0 ...
Page 36: ...2 8 Connections Pinouts 918494 DaqBoard 3000 Series User s Manual ...
Page 40: ...4 2 Calibration 918494 DaqBoard 3000 Series User s Manual ...
Page 80: ...6 12 Setpoint Configuration for Output Control 908794 DaqBoard 3000 Series User s Manual ...
Page 102: ...DBK215 pg A 14 948894 Appendix A ...
Page 104: ...B 2 Hardware Analog Level Trigger 90794 Appendix B ...