You can no w use the software to “configure” your OM-143.
Configuring the device means choosing four key parameters:
•
Sampling time
—How often it makes and stores humidity
and temperature measurements.
•
Temperature unit
—degrees Fahrenheit or Celsius (°F or °C).
•
High and low temperature alarm
setpoints
•
High and low humidity alarm
setpoints
To begin, click on “Set up the USB Data Logger” in the
window shown in the figure above. Doing so will cause the
following window to appear.
Using your mouse, trackball or touchpad, first choose your
Temp Scale and Sample Rate (sampling time). With regard to
sampling time, using the shortest setting of 10 seconds may
result in too much essentially identical data being stored in
memory, potentially limiting its capacity to capture important
events such as an unexpected excursion of temperature or