background image

slowly and carefully to avoid bending. Do not hold the head when
inserting, always hold the stem.

Do not place a "Dial Temp" thermometer in contact with a
material where electrical voltage potential is present. 

SECTION 4 CALIBRATION

OMEGA K-79 Series Pocket Thermometers are factory calibrated. The
alloys used in the bi-metal element determine the span calibration and
are not adjustable. The dial head itself, however, can be reset. This is
done at ambient temperature, where the pointer remains stable, and the
dial is adjusted.
To reset the dial:
1. Determine the room temperature by using a reference thermometer

of known accuracy.

2. Firmly hold the hex nut at the base of the dial with a 7/16” wrench.
3. With your other hand, firmly grip the perimeter of the dial. Slowly

turn the dial head until the reading corresponds to room the
temperature.

SECTION 5 SPECIFICATIONS

Accuracy:

1% of scale

Approx. Response Time
in Water

:

4 seconds for one time constant*

Head:

1" diameter (25 mm)

Dial:

White background with crisp letters, marks
and numbers

Stem:

.142 O.D., 5" long, pointed (3.6 mm x 127 mm)

Reset Nut:

7/16" hex (11 mm)

Pocket Case:

Stainless steel with clip and holder loop

Weight:

.65 ounces (18.5 g)

* “Time Constant” is the time required to reach 63.2% of instantaneous
temperature change.

CAUTION

Summary of Contents for K-79

Page 1: ...to read 1 dial has a pointer that is directly driven by the sensitive bi metal helix coil located in the shaft The 0 142 diameter 5 stainless steel probe has a needle tip for penetrating semi solid materials such as fruits and meats The pointed probe is supplied with a tubular pocket case complete with a pocket clip and holder loop See Figure 1 Figure 1 K 79 Thermometer with Pocket Case POCKET CLI...

Page 2: ...ticular note of any evidence of rough handling in transit Immediately report any damage to the shipping agent The carrier will not honor any claims unless all shipping material is saved for their examination After examining and removing contents save packing material and carton in the event reshipment is necessary SECTION 3 OPERATING INSTRUCTIONS For accurate temperature measurement you must immer...

Page 3: ...he dial 1 Determine the room temperature by using a reference thermometer of known accuracy 2 Firmly hold the hex nut at the base of the dial with a 7 16 wrench 3 With your other hand firmly grip the perimeter of the dial Slowly turn the dial head until the reading corresponds to room the temperature SECTION 5 SPECIFICATIONS Accuracy 1 of scale Approx Response Time in Water 4 seconds for one time ...

Page 4: ... medical application used on humans or misused in any way OMEGA assumes no responsibility as set forth in our basic WARRANTY DISCLAIMER language and additionally purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product s in such a manner RETURN REQUESTS INQUIRIES Direct all warranty and repair requests inquiries to the OME...