HHP680 Manometer

3

Figure 2. Positions of Switch and Button

HHP680 Calibration Procedure

Only qualified technicians who have access to the items, as
following should perform the following calibration procedure.

Equipment:

Pressure Calibrator, Pressure Pump

Pressure Calibration

1. Connect the calibrator. meter and pressure pump.
2. Set the range to the 0.01", then press ”Zero” button.
3. Apply a 19.00 inH

2

O pressure to the meter and calibrator. (do not apply

pressure that exceeds 

20inH

2

O

.)

4. Adjust VR2 until display reads 19.00.

N O T E

2

4

3

1

HHP680

MANOMETER

(   > 2 Sec) in WC

H

0.01"

0.01"

0.1"

0.1"

MAX

ZERO