background image

(10) “TYPE”  button  (MAIN)

Press this button to change the type of thermocouple in

the main display (K/J/T/E/R/S/N). If the inputs of the main

and second displays are the same, then pressing this button

will change the thermocouple type for both displays.

(1

1) “TYPE”  button  (SECOND)

Press this button to change the type of thermocouple in

the second display (K/J/T/E/R/S/N). If the inputs of the

main and second display are the same, then pressing this

button will change the thermocouple type for both displays.

SETUP

   MODE

The following functions can only be used in the setup

mode.

(2) “[Limits]”  button  (Hi/Lo  limit  setting)

 and 

Mini  DIN output

Press the LIMITS button to enter the Hi/Lo Limit setup

function, the “LIMIT”, “HI”, and main display will blink

on the LCD, and the previous settings are displayed. Press

the number button printed on the overlay in white to

set Hi or Lo limit value. the resolution of Hi/Lo limit setup

is 1 count. The button “-” (same button as the ESC) can be

used to set negative values. Setting is from left to right digit.

Press the “ENTER” button to confirm each setting.

3) “[INVT]”  button  (Interval   time   setting)

T

o

 setup the time interval for the log function, press the

[INVT] button to set the time.  The indicator “INV” will be

blinking on the topright of the LCD and the previous settings

will be displayed. Press the number button printed on the 

overlay in white to change the time interval. Setting is from

left to right of the following format (HH:MM:SS). Press the

overlay “ENTER” button to confirm. T

o

 exit this function,

press the ESC button.

HH: Interval Hour (0~23)

MM: Interval Minute (0~59)

SS: Interval Second (0~59)

MAX: 23:59:59

MIN:00:00:01

(8) “[APO]” button (Auto  power

  off /time

 

setting,

min.  5  minutes)

Use this function to change the time for the Auto Power

Of
f (APO) function. Press the [APO] in setup mode, and the

indicator “APO” and the main display will blink on the

LCD, and the previous setting will be displayed. The

default time for APO is 5 minutes. Press the Number button

printed on the overlay in white to set the APO time.

Press the overlay “ENTER” button to confirm. T

o

 exit this

function without changing the setting, press the ESC

button.

MAX: 19999 minutes

MIN:  0005 minutes

(9) “[OFS]”  button  (Thermocouple  offset  adjust)

When the main display input is T1 and socket thermo-

couple is connected, the instrument can adjust the of

fset of

the thermocouple or vice-versa for T2. In the setup mode,

press this button to enter the thermocouple Of

fset Setup

Function (OFS) and the indicator CAL

 should be displayed on

the top righthand corner of the LCD while the main display 

blinks. While the previous setting is displayed, press the 

Number button printed on the overlay in white to change 

or set the of

fset of the thermocouple. The resolution of 

the setup is 0.1. Press the “-” button to set the negative 

 

value. Press the overlay “ENTER” button to confirm.

MAX: 1999.9 C/F

(12) “[TIME]”  button  (System   date   and   time   setting)

T

o

 set or change the system time, press the [TIME]

button in the setup mode. The time and date on the third

display panel should blink. Enter from left to right in the

following format: YY

:MM:DD and HH:MM:SS. Press

the Number button printed on the overlay in white word to

set the system date and time. Press the “ENTER” button

to confirm. Exit this function by pressing the ESC button.

PROCEDURE   OF

   CALIBRA

TION

1. Make ADJ pin short.

2. T2 input for a voltage of 70mV

 will show a value 

higher than 28000 2dgts at the third display and then

press “set[]” key in the next step.

3. T1 input for a temperature of 0C will show a value

5C at the main display

. Press “set[]” key

to complete the calibration.

4. T

ake of

f the short pin.

OPERA

T

OR   MAINTENANCE

Battery Replacement

Power is supplied by a 9 volt “transistor” battery

.

(NEDA

 1604, IEC 6F22). The “

 ” appears on the LCD

display when replacement is needed. T

o

 replace the bat-

tery

, remove the two screws from the back of the meter and

lift of

f the battery cover

. Remove the battery from battery

contacts.

W

ARNING

T

o

 avoid possible electrical shock, disconnect the

thermocouple connectors from the thermometer

before removing the cover

.

W
A

RRANTY

/DISCLAIMER

OMEGA ENGINEERING, INC. warrants this unit to be free of

defects in materials and workmanship for a period of 

13

months

from date of purchase. OMEGA

’s

 

W
ARRANTY 

adds an

additional one (1) month grace period to the normal 

one (1)

year product warranty

to cover ling 

and shipping time.

This ensures that OMEGA

’s

 

customers receive maximum

coverage on each product. 

If the unit malfunctions, it must be returned to the factory for

evaluation. OMEGA

’s

 

Customer Service Department will issue

an Authorized Return (AR) number immediately upon phone or

written request. Upon examination by OMEGA, if the unit is

found to be defective, it will be repaired or replaced at no

charge. OMEGA

’s

 

WARRANTY does not apply to defects

resulting from any action of the purchaser

including but not

limited to mishandling, improper interfacing, operation outside

of design limits, improper repair

or unauthorized modification.

This W

ARRANTY 

is VOID if the unit shows evidence of having

been tampered with or shows evidence of having been

damaged as a result of excessive corrosion; or current, heat,

moisture or vibration; improper specification; misapplication;

misuse or other operating conditions outside of OMEGA

’s

control. 

Components in which wear is not warranted, include but

are not limited to contact points, fuses, and triacs.

OMEGA is pleased to offer suggestions on the use of its

various products. However

OMEGA neither assumes

responsibility for any omissions or errors nor assumes

liability for any damages that result from the use of its

products in accordance with information provided by

OMEGA, either verbal or written. OMEGA warrants only

that the parts manufactured by it will be as specified and

free of defects. OMEGA MAKES NO OTHER

WARRANTIES OR REPRESENT

A

T

IONS 

OF ANY KIND

TITLE, AND ALL IMPLIED W

ARRANTIES 

INCLUDING ANY

W
ARRANTY OF 

MERCHANT

ABILITY 

AND 

FITNESS 

FOR 

A

P

A

R

TICULAR PURPOSE 

ARE 

HEREBY 

DISCLAIMED.

LIMIT

A

TION OF 

LIABILITY

The 

remedies 

of 

purchaser 

set

forth herein are exclusive, and the total liability of

OMEGA with respect to this order


whether based on

contract, warranty

negligence, indemnification, strict

liability or otherwise, shall not exceed the purchase pric

e

of the component upon which liability is based. In no

event shall OMEGA be liable for consequential,

incidental or special damages.

CONDITIONS:  Equipment sold by OMEGA is not intended to be

used, nor shall it be used: (1) as a “Basic Component” under 10

CFR 21 (NRC), used in or with any nuclear installation or activity;

or (2) in medical applications or used on humans. Should any

Product(s) be used in or with any nuclear installation or activity,

medical application, used on humans, or misused in any way

,

OMEGA assumes no responsibility as set forth in our basic

W
ARRANTY

/

D

ISCLAIMER language, and, additionally,

purchaser will indemnify OMEGA and hold OMEGA harmless

from any liability or damage whatsoever arising out of the use

of the Product(s) in such a manner.

RETURN REQUESTS

/

INQUIRIES

Direct all warranty and repair requests/inquiries to the OMEGA

Customer Service Department. BEFORE RETURNING ANY

PRODUCT(S) TO OMEGA, PURCHASER MUST OBT

AIN 

AN

AUTHORIZED RETURN (AR) NUMBER FROM OMEGA

S

CUSTOMER SER

VICE 

DEP

A

R

TMENT 

(IN ORDER TO A

VOI

D

PROCESSING DELA

YS). 

The assigned AR number should then

be marked on the outside of the return package and on any

correspondence.

The purchaser is responsible for shipping charges, freight,

insurance and proper packaging to prevent breakage in transit. 

 

FOR 

W
ARRANTY

RETURNS,

please have the following

information available BEFORE

contacting OMEGA:

1. Purchase Order number under

which the product was

PURCHASED,

2. Model and serial number of th

e

product under warranty

, and

3. Repair instructions and/or

specific problems relative to

the product.

FOR 

NON-W

ARRANT

Y

REP

AIRS,

consult OMEGA for

current repair charges. Have the

following information available

BEFORE contacting OMEGA:

1. 
Purchase Order number to

cover the COST of the repair

,

2. 
Model and serial number of

the product, and

3. 
Repair instructions and/or

specific problems relative to

the product.

OMEGA

’s

 policy is to make running changes, not model

changes, whenever an improvement is possible. This affords our

customers the latest in technology and engineering.

OMEGA is a registered trademark of OMEGA ENGINEERING, INC.

© Copyright 2004 OMEGA ENGINEERING, INC. All right

s

reserved. This document may not be copied, photocopied,

reproduced, translated, or reduced to any electronic medium or

machine-readable form, in whole or in part, without the prior

written consent of OMEGA ENGINEERING, INC.

Where Do I Find Ever

ything I Need 

for Pr

ocess Measurement 

and Contr

ol? 

OM

EGA…Of Course!

Shop online at omega.com

TEMPERA

TURE

䡺⻬

Thermocouple, R

T

D & Thermistor Pr

obes, 

Connectors, Panels & Assemblies

䡺⻬

W

ir

e: Thermocouple, R

T

D & Thermisto

r

䡺⻬

Calibrators & Ice Point Refer

ences

䡺⻬

Recor

ders, Contr

ollers & Pr

ocess Monitors

䡺⻬

Infrar

ed Pyr

o

meters

PRESSURE, STRAIN AND FORCE

䡺⻬

T

ransducers & Strain Gages

䡺⻬

Load Cells & Pr

essur

e Gages

䡺⻬

Displacement T

ransducers

䡺⻬

Instr

umentation & Accessories

FLOW/LEVEL

䡺⻬

Rotameters, Gas Mass Flowmeters & Flow

Computers

䡺⻬

Air V

elocity Indicator

s

䡺⻬

T

u

rbine/Paddlewheel System

s

䡺⻬

T

o

talizers & Batch Contr

ollers

pH/CONDUCTIVITY

䡺⻬

p

H

 Electr

odes, T

esters & Accessories

䡺⻬

Benchtop/Laboratory Meters

䡺⻬

Contr

ollers, Calibrators, Simulators & Pumps

䡺⻬

Industrial p

H

 & Conductivity Equipment

DA
TA

 ACQUISITION

䡺⻬

Data Acquisition & Engineering Softwar

e

䡺⻬

Communications-Based Acquisition Systems

䡺⻬

Plug-in Car

d

s for Apple, IBM & Compatible

s

䡺⻬

Datalogging Systems

䡺⻬

Recor

ders, Printers & Plotters

HEA

TER

S

䡺⻬

Heating Cable

䡺⻬

Cartridge & Strip Heaters

䡺⻬

Immersion & Band Heaters

䡺⻬

Flexible Heaters

䡺⻬

Laboratory Heaters

ENVIRONMENT

A

L

MONITORING AND CONTRO

L

䡺⻬

Metering & Contr

ol Instr

umentatio

n

䡺⻬

Refractometers

䡺⻬

Pumps & T

ubing

䡺⻬

Air
, Soil & W

ater Monitors

䡺⻬

Industrial W

ater & W

astewater T

reatment

䡺⻬

pH, Conductivity & Dissolved Oxygen

Instr

uments

WHA

T

SOEVER, EXPRESSED OR IMPLIED, EXCEPT THA

T

 OF

Reviews: