3
side of the Optical Sensor Switch. Alternatively power can be
connected via the DC power jack on the bottom side of the unit. If
you are using your own power supply, be sure that the voltage
level is b12 and +15 Vdc.
SUPPLYING DC POWER TO THE POWER JACK AND THE "D"
CONNECTOR AT THE SAME TIME WILL DAMAGE THE
METER. DC POWER JACK POLARITY IS CENTER POSITIVE.
The DIP switch, located near 9-pin D-connector can be used to set
up custom settings separately for High and Low alarm.
Figure 3.1
, Optical Sensor Switch 9-Pin "D" Connector Configuration.
WARNING: DO NOT CONNECT POWER SUPPLY WITH MORE
THAN 15Vdc VOLTAGE.
Important notes:
In general, "D" Connector numbering patterns are standardized. There are,
however, some connectors with nonconforming patterns and the
numbering sequence on your mating connector may or may not coincide
with the numbering sequence shown in our pin configuration table above. It
is imperative that you match the appropriate wires in accordance with the
correct sequence regardless of the particular numbers displayed on your
mating connector.
Make sure power is OFF when connecting or disconnecting any cables in
the system.