Calibration

Note: For best results,  calibration should be performed with the 
intended output load, in the operating environment, mounted 
on a DIN rail, allowing at least one hour for thermal equilibrium 
of the system.

1. Install the module on to a piece of DIN rail and the ACPB rail 
mounting combination.  See the ACPB rail Data sheet for details.

Note: An ACPB rail is required to deliver power to the modules.  

See ordering information.

2. Connect the input to a calibrated DC source and the output 
to a voltage or current meter. Apply power and allow the system 
to reach thermal equilibrium (approx. 20 minutes).

3. Adjust the input signal to the desired maximum and observe 
that the green LED is on or flashing.  Push the CAL button and 
hold it down for more than 5 seconds (until the yellow and Green 
LEDs are flashing). 
 

Figure 1: Input Connection

Figure 2: DRI-DC and DRI-SP-AC Calibration Flow Chart

Type

Selector SW1

5

6

7

8

Unipolar

-

-

-

Bipolar

-

-

-

Reverse

-

-

-

Direct

-

-

-

Current

-

-

Voltage

-

-

Key:

 = 1 = ON or Closed; - = n/a

Ranges

Selector SW1

Voltage

Current

1

2

3

4

20mV

2mA

50mV

5mA

100mV

10mA

200mV

20mA

500mV

50mA

1V

100mA

2V

5V

10V

25V

50V

100V

Key:

 = 1 = ON or Closed

Ranges

Selector SW2

Output

1

2

3

4

5

6

7

8

0 to 5V

0 to 10V

0 to 1mA

4 to 20mA

0 to 20mA

Key:

 = 1 = ON or Closed

Table 1: Input Range Settings for Single Channel Module 

(DRI-DC)

Table 3: Output Range Settings for Single Channel Module 

(DRI-DC)

Table 2: Input Function Settings for Single Channel Module 

(DRI-DC)

Table 4: I/O Range & Function Settings 

for Dual Channel Module (DRI-SP-AC)

Figure 1: Input Connection 

4. When the yellow and green LEDs stop flashing, the yellow 
and red LEDs will be on.  Push the CAL button momentarily (the 
yellow and green LEDs will be on).

Note: To quit the calibration mode and reset the unit, push the 
CAL button and hold for more than 5 seconds. Or, wait for more 
than two minutes and the unit will timeout and  reset itself to the 
previously stored  calibration.

5. Apply the maximum input signal level desired, and push the 
CAL button to store  (the yellow LED will be on).

6. Apply the minimum input signal level desired, and push the 
CAL button to store  (the green and red LED will be on.

7. Adjust the input signal while monitoring the output signal until 
the output is at the desired maximum level (e.g. 20.00mA), then 
push the CAL button to store (the red LED will be on).

8. Adjust the input signal while monitoring the output signal until 
the output is at the desired minimum level (e.g. 4.00mA), then 
push the CAL button to store (the yellow, green and red LEDs 
will be on). 

9. Push the CAL button one final time to store the calibration data. 
The green LED will be on if the input is within the calibrated range. 