8-34
STRAIN
0
8
Scaling to
Display
Engineering
Units
MENU
SUBMENU 1 SUBMENU 2 DESCRIPTION
INP.3=0
Unipolar input
(S2B jumper omitted
INP.4=0
Std, for current inputs
INP.5=
Disregard not used
INP.6=0
Disables IN.SC.OF
INP.7=0
Non ratiometric input
IN.SC.OF
Skip past
DEC PT
FFFFFF
No decimal point for
calibration run
(Desired decimal point
will be entered later)
CNT BY
Press the ‘RESET’ button
two times to return to
the run mode
2. Now you are in the run mode and the transducer is connected to
the meter. Simulate two loads on the transducer and record the
meter’s display as “INPUT1” and “INPUT2” and record what
each of these readings should display (“READ 1” and
“READ 2”) on a piece of scrap paper or in the spaces provided
below. These numbers will be the INPUT values entered in the
“IN.SC.OF” scaling. Normally the two points that are calibrated
are the no load condition and 3/4 full scale. However, you may
use any two measurements.
“INPUT1”
“INPUT2”
“READ 1”
“READ 2”
Assume that you simulate no load and the display indicates “19590”,
and when you simulate 4000 PSI the display indicates “80150”.
The following data will now be entered into the 2-point scaling
“IN.SC.OF” method.
“INPUT1”:
019590
“READ 1”:
0000
“INPUT2”:
080150
“READ 2”:
4000