background image

Using This Quick Start Manual

Use this Quick Start Manual to get your High Performance
Process Indicator up and running.  These instructions use
the factory default settings of current input and 

24 Vdc

sensor excitation. 

The latest complete Communication and Operational
Manual as well as 

free

Software are available at

www.

omega

.com

or on the

CD-ROM

enclosed with your shipment

.

To start your unit you:
• connect ac power
• wire the sensor
• configure the meter, using the front panel buttons and the

configuration menus

Before You Begin

Check the serial label on the top of the meter case.  The
model number’s last 3 digits indicate if the meter is set up
for voltage or current.  C1 - C2 are current ranges and 
DC1 - DC10 are voltage ranges.  

This quick start manual

uses the C2 (4 - 20mA) range

.  If your meter is setup for a

range other than C2, consult the Operator’s manual.

In addition to the unit and related parts, you will need the
following items to set up your unit:
• ac power, as listed on meter’s Product/ID
• 4-20mA input (eg: load cell, calibrator)
• 1/8” Phillips head screwdriver
• 1/8” flat blade screwdriver

Warning: Do not connect AC power to your
device until you have completed all input and
output connections. This device must only be

installed by a specially trained electrician with
corresponding qualifications. Failure to follow all
instructions and warnings may result in injury!

Connect ac Power

1. Remove the rear protective cover and set it aside.

The cover is secured with a Phillips-head screw.

2. Locate connector P1 on the bottom-left-rear of the unit.

The connector has three screw-down terminals.

3. Insert the correct wire in each terminal and tighten the

lockdown screw.  Tug gently on each wire to verify the
connection.

1
2
3

P1

N

~AC LINE
~AC NEUTRAL

PROTECTIVE EARTH GND

L

L

FUSE

N

SWITCH

GREEN WIRE

EARTH

LINE

NEUTRAL

Check for proper Earth grounding
in the power distribution system (single phase).

Wiring a Current Transmitter

Follow these steps to wire a current input sensor with
sensor excitation.
1. Locate connectors P3 & P9 on the right-side-rear of the unit.

2. Attach the wires and tighten the retaining screws.  Tug

gently on the wires to verify the connection.

Wiring Example:  Current Input with Sensor Excitation

3. Apply ac power. The front panel of the unit flashes RESET2.

If it does not:
a. Remove ac power.
b. Verify the P1 power and P3 and P9 input connections.
c. Check your power source.
d. Check your signal source.

4. Replace the rear cover.  Thread the sensor wires through

the slots on the side of the cover.  Replace the rear cover
screw.

5. Apply ac power once again.

Configure the Meter

Use the front panel buttons to access the configuration
menus, to either verify or set the unit values. The function
of each button is described below:

Meter Button Descriptions

Press This Button To:

sequentailly recall previous setpoint 
settings (run mode only), store new
setpoint values.

access the configuration program
menus and move from one menu
to the next.

enter and scroll through a submenu.

change the value of a submenu.

move backward one menu (press
once) or exit the configuration
menus (press twice).  

N/C

+S

–S

P9

1

2

3

1

2

3

P3

N/C

RTN

(+)

(

-

)

(4-20mA)

+E

–E

JUMPER
USER
PROVIDED

Quick Start Setup

Quick Start Setup scales your meter to read 0 to 100.0 for a 
4 - 20mA input.  This setup procedure is actually a series of
short tasks.  To complete Quick Start Setup successfully, it
is important to perform all tasks in the order outlined in the
Quick Start Setup. 

Select Input Range

1. Press MENU until the meter displays:

2. Press  

/MIN until the meter flashes:

Press MENU and “CURRNT” stops
flashing.
Press  

/MIN to display a flashing 

input range of 4-20mA or 0-20 mA.
Press 

/MAX until the meter displays

“4-20mA.”

3. Press MENU. The meter stores changes (if applicable),

then displays “RDG.CNF”:

Set Up Reading Configuration

1. Press 

/MIN to display “RDG.1”. 

Press 

/MAX until the meter displays

“RDG.1=0”. Selecting “RDG.1=0”
enables direct formatting.

2. Press 

/MIN to display “RDG.2”.

Press 

/MAX until the meter displays 

“RDG.2=1”. Selecting “RDG.2=1”
enables an active decimal point.

3. Press MENU.  The meter stores changes (if applicable),

then displays “RDG SC”:

Set Reading Scale Factor and Reading Offset

1. Press  

/MIN to display the scale factor

value.  Change the value, if necessary,
to read “1.00000”  (Use 

/MIN to scroll

through digits and 

/MAX to change

the digit’s value).

2. Press MENU.  The meter stores changes (if applicable)

then displays “RDG.OF”:

3. Press 

/MIN to display the offset value.

Change the value, if necessary, to read
“000000”  (Use 

/MIN to scroll

through digits and 

/MAX to change

the digit’s value). 

4. Press MENU.  The meter stores change,

if applicable, then displays “IN CNF”.

Set Default Input Scale and Offset

1. Press 

/MIN until the meter flashes.

Change the value, if necessary, to read 
“INP.6=0”  (Use  

/MIN to scroll 

through  INP submenu choices and 

/MAX to select 0 as the value). Setting

“INP.6=0” disables input scale and offset.

2. Press MENU.  The meter stores changes, if applicable,

then displays “IN.SC.OF”.

2

3

4

START HERE

/MIN

/MAX

RESET

MENU

SETPTS

1.00000

SETPTS

/MAX

/MIN

MENU

RESET

stored

SETPTS

/MAX

/MIN

MENU

RESET

IN.SC.OF

SETPTS

/MAX

/MIN

MENU

RESET

stored

SETPTS

/MAX

/MIN

MENU

RESET

 rdg.of

SETPTS

/MAX

/MIN

MENU

RESET

000000

SETPTS

/MAX

/MIN

MENU

RESET

stored

SETPTS

/MAX

/MIN

MENU

RESET

 CNTBY

SETPTS

/MAX

/MIN

MENU

RESET

INp.6=0

SETPTS

/MAX

/MIN

MENU

RESET

stored

SETPTS

/MAX

/MIN

MENU

RESET

rdg sc

SETPTS

/MAX

/MIN

MENU

RESET

 INPUT

SETPTS

/MAX

/MIN

MENU

RESET

CURRNT

SETPTS

/MAX

/MIN

MENU

RESET

RDG.2=1

SETPTS

/MAX

/MIN

MENU

RESET

stored

SETPTS

/MAX

/MIN

MENU

RESET

rdg.CNF

SETPTS

/MAX

/MIN

MENU

RESET

RDG.1=0

SETPTS

/MAX

/MIN

MENU

RESET

Reviews: