3. Additional documentation

1. Front view

1.

 Open the instrument (see section 7).

2.

 Select the jumpers for the desired signal range (see section 8).

3.

 Close the instrument (see section 7).

4.

 Connect the signal and the power (see section 6).

5.

 Configure the instrument from the ‘Configuration menu’ (see section 9).

6.

 If you need additional information, see section 3.

5. Installation and start-up

6. Connections

7. How to open the instrument

DP20

Model

-

Option 1

-A1  (1 relay)

-M1  (analog output)

-S1  (Modbus RTU)

-(empty)

-

Option 2

-A2  (1 relay)

-(empty)

4. How to order

2. Dimensions and panel cut-out (mm)

8. Internal structure and jumpers for range selection

MQS5802_0618

Key ‘LE’

Key ‘SQ’

1. Exit the menu

Key ‘UP’

1. Enter into ‘Fast access’ 

2. Next value

Units

Alarm 1 and 2

Fuse

Option M1 (4/20 mA)

Option A1 (relay)

Protection fuse value :

  250 mA time-lag for power voltage > 50 Vac/dc 

  400 mA time-lag for power voltage < 50 Vac/dc

A B C

Common

Vexc.

4/20 mA

A B C

NC

Com.

NO

Option S1 (Modbus RTU)

A B C

Common

B

A

Option 2

Option 1

1 2 3 4 5

8 9 0

-Vdc  / Neutral

+Vdc  / Phase

 

  Not Connected

A B C

Option A2 (relay)

NC

Com.

NO

G I

H

G I

H

~

Vac, ±Vdc, resistance, mA, pot 

therm, Pt+, Ni+, PTC+, NTC+

~

600 Vac, 

~

200 Vac  

±600 Vdc, ±200 Vdc

~

5 Aac, ±5 Adc

neutral, 0 V, common

Vexc, Pt100 3 wires, pot+ 

external control

NTC probes ‘R

25

’ 

(configurable)*

Jumpers 

‘S’

Jumper 

‘T’

Range  

of measure

Accuracy 

(% of reading)

Beta 

(configurable)

10K

F K

4-5

-60

º

C to

150

º

C

<1.5 % of 

reading

3500

*‘Beta’ configurable (2000 to 5500). R

25

 configurable. Resistance measure from 100R to 1MR.

AC ranges

Scalable

Jumpers 

‘S’

Jumper 

‘T’

Accuracy 

(% FS)

~ 

600 Vac

from 9999  

to -1999

G I

4-5

<0.30 % 

(up to 150 Hz)

~ 

200 Vac

I

~ 

20 Vac

A I

~ 

2 Vac

B I

~ 

200 mVac

C I

~ 

60 mVac

E I

~ 

5 Aac

I

<0.50 % 

(up to 150 Hz)

~ 

20 mAac

D I

DC ranges

Scalable

Jumpers 

‘S’

Jumper 

‘T’

Accuracy 

(% FS)

±600 Vdc

from 9999  

to -1999

G

4-5

<0.20 %

±200 Vdc

---

±20 Vdc

A

±2 Vdc

B

±200 mVdc

C

±60 mVdc

E

<0.25 %

±5 Adc

- - -

±20 mAdc

D

<0.15 %

PTC probes  

Family

Jumpers 

‘S’

Jumper 

‘T’

Range in ºC 

(in ºF)

Total error

KTY-121

F

4-5

-55 / 150 ºC 

(-67 / 302 ºF)

<1 º

KTY-210

F H K

KTY-220

F H K

Pt and Ni 

probes

Jumpers 

‘S’

Jumper 

‘T’

Range in ºC 

(in ºF)

Total 

error

Current at 

sensor

Pt100  

(3 wires)

F H J

5-6

-200 / 700 ºC 

(-328 / 1292 ºF)

<1º

< 900 uA

Pt100  

(2 wires)

F H

4-5

-200 / 700 ºC 

(-328 / 1292 ºF)

< 900 uA

Pt500

F

-150 / 630 ºC 

(-238 / 1166 ºF)

< 90 uA

Pt1000

F

-190 / 630 ºC 

(-310 / 1166 ºF)

< 90 uA

Ni100

F H

-60 / 180 ºC 

(-76 / 356 ºF)

< 900 uA

Ni200

F H

-60 / 120 ºC 

(-76 / 248 ºF)

< 900 uA

Ni1000

F

-60 / 180 ºC 

(-76 / 356 ºF)

< 90 uA

Thermocouples Jumpers 

‘S’

Jumper 

‘T’

Range in ºC 

(in ºF)

Total error 

(cold junction included)

tc. K

E

4-5

-100 / 1350 ºC 

(-148 / 2462 ºF)

<3 º 

tc. J

-100 / 1200 ºC 

(-148 / 2192 ºF)

tc. E

-100 / 1000 ºC 

(-148 / 1832 ºF)

tc. N

-100 / 1300 ºC 

(-148 / 2372 ºF)

tc. L

-100 / 900 ºC 

(-148 / 1652 ºF)

tc. R

E J

0 / 1768 ºC 

(32 / 3214 ºF)

tc. S

0 / 1768 ºC 

(32 / 3214 ºF)

tc. T

-100 / 400 ºC 

(-148 / 752 ºF)

tc. C

E

0 / 2300 ºC 

(32 / 4172 ºF)

<5 º

tc. B

E J

700 / 1820 ºC 

(1292 / 3308 ºF)

Resistance  

ranges

Scalable

Jumpers 

‘S’

Jumper 

‘T’

Accuracy  

(% reading)

0 to 5 K

from 9999  

to -1999

F H K

4-5

<1.5 % of 

reading

0 to50 K

F K

Process  

signals

Scalable

Jumpers 

‘S’

Jumper 

‘T’

Accuracy 

(% FS)

4/20 mA

from 9999  

to -1999

D

1-2

*

<0.15 %

0/10 Vdc

A

<0.20 %

* Place jumper ‘T’ at position 1-2 for +15 Vdc excitation voltage at terminal 5. Optionally, 

place jumper ‘T’ at position 4-5 to work with ‘external contact’ at terminal 5. 

Potentiometers 

nominal value

Scalable

Jumpers 

‘S’

Jumper 

‘T’

Accuracy  

(% FS)

500 R to 20 K

from 9999  

to -1999

A

2-3

<0.5 %

!

(active  output  A-B) 

(passive output B-C)

Risk of electric shock. Removing the front cover will grant access to internal 

circuits which may be at dangerous voltage. Disconnect the input signal and 

the power supply to prevent electric shock to the operator. Operation must be 

performed by qualified personnel only.

Use a flat screwdriver to unlock clips ‘D’, C’, ‘B’ 

and  ‘A’,  in  this  order.  Remove  the  front  filter. 

Gently  let  the  internal  boards  slide  out  of  the 

instrument.
To reinsert the boards in the housing :

1.

 make sure  that the boards are correctly con-

nected to the displays pins 

2.

 slide the boards into the housing guides 

3.

 place the front filter at corner X, and then 

insert clips ‘A’, ‘B’, ‘C’ and ‘D’ in this order.

X

A

C

B

D

Internal Jumpers

Opt.1

Opt.2

Opt.1

Display module

Optional modules

Housing

Front filter

Input signal terminal

Jumpers ‘S’  

for signal range selection

Displays

F
E

G

H

I

J

D

B

C

A

K

Jumpers  not  used  can  be 

stored in the following three 

‘no contact’ positions.

F
E

G

H

I

J

D

B

C

A

K

6

5

4

3

2

1

Jumper ‘T’  

for terminal 5 function selection

DP20  shown  smaller 

than actual size

Quick installation guide DP20 (page 1/2)

www.omega.com/ DP20.html

Detail of the female 

 

terminal provided.

96 mm

48 mm

91 mm

(including terminal)

16

8

75

(0.63)

(0.31)

(2.95)

Panel 

cut-out

92

(3.63)

44

(1.74)

(3.78)

(1.89)

(3.58)

TM

Frequency

Scalable

Jumpers ‘S’

Jumpers ‘T’

Accuracy 

(% of reading)

15 Hz to 100 Hz

from 9999 

to -1999 

select Vac or 

Aac range

4-5

<0.15%  

of reading 