background image

An  O ME GA Te c h n o l o g i e s  Co m p a n y

 M2106/1294

INSTRUCTION

SHEET

DP18 Series

Models DP18-RV1 / - RV2 / - RV3 / - RV4
True RMS Voltmeter

CCD1294APC2BA

1

10

D
C
B
A

+

-

REAR VIEW

Setpoint R1
(option)

Setpoint R2
(option)

FRONT VIEW

FIG. 2

FIG. 1

SIGNAL
INPUT

SPAN : ADJUST THE MAXIMUM READING.
OFFSET : NOT IN SERVICE

SPAN ADJUSTMENT IS ACCESSIBLE BEHIND THE LOW DOOR.

Isolated digital output

(option)

HOLD,

 latch the reading

display, while these pins
are linked

ANALOG OUTPUT
(OPTION)

POWER SUPPLY

see instrument label

DISPLAYS . . . . . . . . . . . . . . . . . . . . . .  3      digits, 7 segment, red LED
HEIGHT . . . . . . . . . . . . . . . . . . . . . . . .  0.56" (14.2 mm.)
SYMBOLS . . . . . . . . . . . . . . . . . . . . . .  ± 1.8.8.8
DECIMAL POINTS . . . . . . . . . . . . . . . .  Selectable by jumpers
OVERRANGE INDICATION . . . . . . . . .  Display flashes 1999

INPUT CONFIGURATION . . . . . . . . . . .  Bipolar single
ZERO . . . . . . . . . . . . . . . . . . . . . . . . .  Automatic
SPAN ADJUSTMENT . . . . . . . . . . . . . .  ± 5 %

CONVERTER . . . . . . . . . . . . . . . . . . . .  Dual-slope, average value
POLARITY . . . . . . . . . . . . . . . . . . . . . .  Automatic  ±
SIGNAL INTEGRATION PERIOD . . . . .  80 ms.
READ RATE . . . . . . . . . . . . . . . . . . . .  3.12 / sec.
RANGES . . . . . . . . . . . . . . . . . . . . . . .  See table selection
HOLD . . . . . . . . . . . . . . . . . . . . . . . . .  Optional

NOISE REJECTION NMR . . . . . . . . . . .  50 dB. 50/60 Hz.
ACCURACY . . . . . . . . . . . . . . . . . . . .  0.2 %   ± 1 count
WARMUP TO RATED ACCURACY .  .  .  2  minutes
SPAN TEMPCO . . . . . . . . . . . . . . . . . .  100 ppm.
ZERO TEMPCO . . . . . . . . . . . . . . . . . .  2.5 µV/°C.

OPERATING TEMPERATURE . . . . . . .  0 to 50 °C (32 to 122 °F)
STORAGE TEMPERATURE . . . . . . . . .  -40 to 80 °C (-40 to 176 °F)
WEIGHT . . . . . . . . . . . . . . . . . . . . . . . .  310 gr. (0.68 lb)
CASE MATERIAL . . . . . . . . . . . . . . . .  ABS, DIN 43700, black color
STANDARD POWER . . . . . . . . . . . . . .  115 Vac. ± 10 %, 50/60 Hz.
POWER CONSUMPTION . . . . . . . . . . .  5.5 VA for AC. / 3.5 W for DC
ELECTRICAL CONNECTIONS . . . . . . .  Push-in cable connectors
BURN-IN . . . . . . . . . . . . . . . . . . . . . . .  24 h.

½

SPECIFICATIONS

SPAN

OFFSET

FIG. 3

RANGE SELECTION

H    J     L

IC2

IC1

200 mVac
     2 Vac

È

È

Ç

Ç

CONNECT SIGNAL INPUT, BETWEEN  A-B or A-C
IN FUNCTION OF THE  VOLTAGE INPUT

CLOSE THE APPROPRIATE DECIMAL POINT JUMPER (FIG.5), CORRESPONDING TO THE RANGE SELECTED

PUT THE S1 SELECTOR PLUG-IN JUMPER, IN THE RANGE DESIRED.

K

200 mV

DISPLAY
READING

RESOLUTION

   mV.

            CLOSE JUMPER
  AMPLIFIER      DECIMAL POINT
       FIG. 3                   FIG. 5

IMPEDANCE
   

 

   Ω

D3

D3

-----

D2

D1

RANGE
   Vac.

A-B

A-C

DP18-RV4

DP18-RV1

DP18-RV2

DP18-RV3

MODEL

S1 SELECTOR

TRUE-RMS INSTRUMENT 3½ DIGITS, WITH FIVE SELECTABLE RANGES ON THE SIGNAL CONDITIONER  BY PLUG-IN JUMPERS. THE
READING CAN NOT BE MODIFIED IN ENGINEERING UNITS,UNLESS THE AMPLIFIER IS ALSO MODIFIED. THEREFORE USING THE
STANDARD AMPLIFIER THE DISPLAY WILL SHOW EXACTLY THE VALUE PRESENT AT THE INPUT SIGNAL.

A

B

C

SIGNAL CONDITIONER
 COMPONENTS VIEW

I     K

   2 V

 20 V

 200 V

500 V

K,L

H,L

I,L

J,L

  1

10

100

1000

 

 

 0.1

10 G

3.3 M

3 M

1.999 V

19.99 V

199.9 V

500 V

1.999 mV

 20 Vac

200 Vac
500 Vac

Reviews: