Introduction 

1

 

INTRODUCTION 

The Platinum

 Series Universal Benchtop Digital Controller, is ideal for laboratory and other applications 

requiring portable, temperature, process or strain, measurement and control.  It features a universal 
input which reads most temperature, process and bridge type inputs. The Benchtop Digital Controller 
has excellent accuracy and is factory calibrated to give optimum performance over its full operating 
range.  

1.1

 

Safety and Precautions 

It is important to read and follow all precautions and instructions in this manual and other referenced 
manuals, before operating or commissioning this device, as it contains important information relating to 
safety and EMC. 

 

Do not exceed the voltage rating. 

 

Always disconnect the power before changing the signal and power connections. 

 

Do not operate in flammable or explosive environments.   

 

Never operate with a power cord that is not properly rated for use with this unit. 

 

Remove and or disconnect main power cord before attempting any maintenance or fuse 
replacement. 

 

Do not connect and/or operate this unit to a non-grounded or non-polarized outlet or power 
source. 

 

There are no user serviceable parts inside the unit. Attempting to repair or service 
the unit may void the warranty. 
This product is not designed for medical applications.  

1.2

 

Cautions and IEC Symbols 

This device is marked with international safety and hazard symbols shown in the table below, in 
accordance with 2014/35/EU Low Voltage Directive. It is important to read and follow all precautions 
and instructions in this manual before operating or commissioning this device as it contains important 
information relating to safety and EMC. Failure to follow all safety precautions may result in injury 
and/or damage to the controller. Use of this device in a manner not specified by the manufacturer may 
impair protection devices and safety features provided by the unit. 
 

IEC Symbol 

Description 

 

CAUTION

, risk of electrical shock 

 

CAUTION

, refer to accompanying documents 

1.3

 

Statement on CE Marking 

OMEGA

’s

 policy is to comply with all worldwide safety and EMI/EMC regulations that apply to CE 

Certification standards, including 

EMC Directive

 2014/30/EU Low Voltage Directive (Safety) Directive 

2014/35/EU, and 

EEE RoHS II Directive

 2011/65/EU. OMEGA is constantly pursuing certification of its 

products to the European New Approach Directives. OMEGA will add the marking to every applicable 
device upon verification of compliance. 

1.4

 

Available Models 

Model 

Features 

CS8DPT-C24-EIP-A 

Benchtop Controller with 4-Digit Display, Embedded Ethernet, 
Serial Communication, and Isolated Analog Output 

-EIP 

Ethernet 

-C24 

Isolated RS232 and RS485 

-A 

Isolated Analog Output 

CS8DPT 

Benchtop Controller, Universal input with 4-Digit Display 

CS8EPT 

Benchtop Controller, Universal input with 6-Digit Display 

CS8EPT-C24-EIP-A 

Benchtop Controller with 6-Digit Display, Embedded Ethernet, 
Serial Communication, and Isolated Analog Output 

 

Summary of Contents for CS8DPT

Page 1: ...ervice 1 800 622 2378 USA Canada only Engineering Service 1 800 872 9436 USA Canada only Tel 203 359 1660 Fax 203 359 7700 e mail info omega com For Other Locations Visit omega com worldwide The infor...

Page 2: ...RATION AND PROGRAMMING 9 4 1 PLATINUM Series Navigation 9 4 2 Selecting an Input INIt INPt 10 4 3 Set the Setpoint 1 Value PRoG SP1 10 4 4 Set up the Control Output 11 4 4 1 Select an Output Channel P...

Page 3: ...onal safety and hazard symbols shown in the table below in accordance with 2014 35 EU Low Voltage Directive It is important to read and follow all precautions and instructions in this manual before op...

Page 4: ...the packing list it is important to verify all equipment shipped has been delivered as shown in Figure 1 and Table 1 If there are any questions about the shipment please email or call the Customer Se...

Page 5: ...Power Cord DM 230V 16A USA Canada Mexico Power Cord Molded 120V Italy Power Cord IT 230V 16A Continental Europe Power Cord E 10A 240V 10A Europe Power Cord E 16A 240V 16A 3 HARDWARE SETUP This section...

Page 6: ...l communications transmit Table 5 summarizes the universal input pin assignments for different sensor inputs All sensor selections are firmware controlled and no jumper settings are required when swit...

Page 7: ...rs R1 and R2 included in the provided Wire Kit across terminals 4 and 6 and terminals 1 and 8 respectively This allows the Bridge voltage to be measured When powering a bridge from the unit use an int...

Page 8: ...s Output 2 3 Ethernet Port RJ45 10 100Base T Optional 4 AC Main Input Plug IEC60320 C13 Power Socket 90 to 240 Vac 50 60 Hz 5 Output 1 Relay Output 90 240 VAC 3A Max 6 Output 2 SSR Output 90 240 VAC 5...

Page 9: ...es across a set of menu choices at a given level When changing numerical settings press the LEFT button to make the next digit one digit to the left active The RIGHT button moves across a set of menu...

Page 10: ...all process ranges MANL Rd 1 ____ Low display reading IN 1 ____ Manual input for Rd 1 Rd 2 ____ High display reading IN 2 ____ Manual input for Rd 2 LIVE Rd 1 ____ Low display reading IN 1 ____ Live R...

Page 11: ...ng Parameters oFF Turn off the output channel factory default PId Set the output to Proportional Integral Derivative PID Control oN oF Set the output to On Off Control Mode RtRN Set up the output for...

Page 12: ...tion ACtN parameter PRoG PID S ACtn as detailed below Setting Description RVRS Reverse Output increases the process value dRCt Direct Output decreases the process value 5 Set the Autotune Timeout A to...

Page 13: ...10 Volts factory default 0 5 0 to 5 Volts 0 20 0 to 20 mA 4 20 4 to 20 mA 0 24 0 to 24 mA 4 7 2 Set Mode to Retransmission Enable the output by setting the mode to Retransmission PRoG IAN 1 Mode RtRN...

Page 14: ...r electrical equipment for measurement control and laboratory EMC Measurement Category I Category I includes measurements performed on circuits not directly connected to the Mains Supply power Maximum...

Page 15: ...er calibration is available with adjustable gain and offset as well as ice point calibration Refer to the Platinum Series User s Manual M5451 for additional information on user calibration options Opt...

Page 16: ...ation misapplication misuse or other operating conditions outside of OMEGA s control Components in which wear is not warranted include but are not limited to contact points fuses and triacs OMEGA is p...

Reviews: