maintenance for all your surface sensors. A large 

diameter, mirror like, high grade Aluminum allow 

hotplate design, allows the most stability while 

maintaining quick, uniform heat-up on the entire 

system.

Mounting

Place this unit on a bench or table top in horizontal 

position. To operate, be sure to stay away from 

constantly open and close doors, air conditioning vents, 

adjacent computer fans, room air moving fans, open 

windows, etc.

Ambient Temperature

The CL1600 Probe Tester can generate any temperature 

within the specified temperature range when operated 

at ambient temperatures of 5 to 35°C (41 to 95°F).

Top, Front, & Rear Panel Controls and Indicators

ARTWORK/PRODUCT ART/

 DWGS/START HERE ARROW

START HERE

ARTWORK/PRODUCT ART/

 DWGS/START HERE ARROW

2

ARTWORK/PRODUCT ART/

 DWGS/START HERE ARROW

3

ARTWORK/PRODUCT ART/

 DWGS/START HERE ARROW

4

Using This Quick Start Manual

Use this Quick Start Manual with your CL1600 Surface 

Probe Tester for easy installation and basic operation. 

For detailed information, refer to the User’s Guide 

(Manual # M4593).

PRECAUTIONS:

To avoid personal injury please follow all safety 

precautions and operating instructions outlined in 

this section.

•  Do not touch the hot plate or its surrounding 

components at any time; the temperature on 

the hot plate is the same as the temperature 

indicated on the controller display. Please be 

careful.

•  The CL1600 can operate at very high 

temperatures, when testing and handling surface 

sensors, they too will get very hot at the tip end, 

so be sure to have a heat resistance plate to place 

the sensors after testing to allow them to cool 

down.

•  Never turn the power off when the hot plate 

is set to temperatures higher than 100 degrees 

C. Damage to the electronic controller or to the 

heater could occur.

•  Never leave your tester unattended when in use.

• Keep out of reach of all children.

•  Never permanently place or attach any object 

within 6 inches of the plate.

•  Do not operate in explosive environments.

•  Never operate this unit with a power cord other 

than the one provided with this system.

•  This instrument is intended for indoor use only.

•  On initial use, or if this unit hasn’t been powered 

for a period of 10 days, it is recommended to 

turn the unit on and allow to run at 50 degrees C 

for 1 hour, before using at higher settings.

•  Components and heater lifetime can be shorted 

by continuous high temperature use.

•  Never place any foreign materials on the surface 

of the hot plate, fluids can leak into the electronic 

cabinet and into the heater assembly causing risk 

of injury and permanent damage to the tester.

General Information

Omega’s CL1600 Surface Probe Tester is a system 

designed to check and verify surface probe 

readings across a wide range in temperatures, 25 

to 450°C (77 to 842°F). The CL1600 allows periodic 

in-house testing and provides preventive

Process Temperature

This field displays the current temperature of the target 

plate. (red display)

Setpoint Temperature

This field displays the desired calibration block 

temperature. Once the block reaches this desired 

temperature, both displays will read the same value. 

(green display).

Parameter/Access Key:  

Press to scroll through menu parameters.   

  

Raise Key: Used to increase values (hold for fast-step 

progression).

  Lower Key: Used to decrease values (hold for fast-step 

progression).

    Mode Key: This key is inactive. 

Press to save settings and exit a menu level.

AC Power Input:  

The customer connects the power cord to the AC 

Power Input. As a safety precaution, the power cord 

cannot be connected if the fuse compartment is open.
Reference Probe Port:  

The reference probe port enables the user to monitor 

the target plate temperature with an external 

instrument.
Serial Port: 

The female phone jack port allows the customer to 

make RS232 interface with the CL1600.

Changing the Temperature Setpoint

The internal parameters in the microprocessor have 

been set to be functional between 25 and 450°C (77 to 

842°F).

Manual Temperature Change:

To change the temperature set point to any value 

within the allowable range, use the up or down arrow 

keys as needed, holding the corresponding key down 

will allow fast-step progression.

Automated Temperature Change

The CL1600 includes a user interface computer

program, which can be accessed and activated 

through the RS232 port. For a complete description 

of included features, please refer to section 2-5 in 

the User’s Guide.
Process Temperature  

This field displays the current temperature of the 

calibration block.
Setpoint Temperature  

This field displays the desired calibration block 

temperature. Once the block reaches this desired 

temperature, both displays will read the same 

value.

Heat-Up/Cool-Down Cycle Transition Times

 

The following table shows the amount of time 

required to change the target plate temperature 

from one setpoint to another. The starting 

temperature is shown along the left side. The final 

temperature is shown along the top. All transition 

times are approximate.

Cool Down Transition Time

Cool Down Transition Time 

The CL1600 will take 90 minutes to cool down 

from 450°C. Be sure the unit has cooled down and 

it is safe to touch before handling the unit.

RS232 Communication

The RS232 communications port allows 

bi-directional data transfer via a three conductor 

cable consisting of signal ground, receive input, 

and transmit output. It is recommended that 

less than fifty feet of cable be used between the 

computer and this instrument. Note that multiple 

instruments cannot be tied to the same port in this 

configuration. The RS232 port is optically isolated 

to eliminate ground loop problems.

Pinout for A 9-pin PC Serial Port

CL1600

SURFACE PROBE

CALIBRATOR

MADE IN U.S.A.  

90-250V

AC  50/60 Hz, 925W

 

FUSE 2 EA: F 5A

 250V 

RS232 

MODEL NO: 

SERIAL NO: 

OMEGA ENGINEERING INC. 
STAMFORD, CT. 06907-0047 

1-800-826-6342  www.omega.com 

CL1600 

CAUTION: FOR CONTINUED PROTECTION AGAINST 
THE RISK OF FIRE REPLACE WITH ONLY THE SAME 

TYPE, RATING AND SIZE OF FUSE. DISCONNECT

POWER FROM SOURCE BEFORE CHANGING FUSE. 

O    l 

CAUTION!

HOT SURFACE

CAUTION!

HOT SURFACE

SURFACE

PLATE

PROCESS

TEMP.

SETPOINT

TEMP.

PARAMETER

ACCESS KEY

RAISE KEY

RESET

LOWER KEY

MODE KEY

POWER

SWITCH

RS232 PORT

FUSE BOX

AC POWER

INPUT

MODEL/SERIAL

NUMBERS

SAFETY

WARNING

TOP VIEW

FRONT VIEW

REAR VIEW

HEAT

SHIELD

REFERENCE

PROBE

ENTRY

PROBE NOT

SUPPLIED

2.0 min. 4.0 min. 2.0 min. 8.0 min. 11.0 min.

50°C

100°C

200°C

300°C

400°C

450°C

25°C

FROM

TO

16.0 min.

TO: COMPUTER SERIAL PORT

TO: INSTRUMENT

1 2 3 4

TX

RX GND

TX

RX

GND

5

6 7 8 9