Target Plate Air Purge

The BB701 comes with a nitrogen air purge collar 

that surrounds the target plate. When connected to 

a nitrogen gas source the target plate is kept free of 

condensation at lower temperatures. Snow is likely to 

form on the plate when it is kept at temperatures below 

freezing temperature. Snow or condensation will change 

a target’s emissivity. The can lead to inaccuracies unless 

the infrared pyrometer is corrected for this emissivity 

change. It is therefore recommended that the user take 

advantage of this feature. 

Use dry gas or Nitrogen with a dew point below -20°C 

(-4.0°F). The rate of gas inlet should be 0.25 to 0.50 m

2

/

hr. (5 to 10 ft

3

/hr.). Do not exceed the maximum gas 

inlet rate. The port will accept any flexible tubing (vinyl, 

rubber, Teflon, latex) with a ¼" outer diameter, such as 

OMEGA

®

 #TYVY-1418-100-A.

Reference Probe

An RTD probe has been embedded into the target plate 

heater assembly to be used as a reference to check the 

calibrator’s calibration and accuracy. A connector has 

been provided with your calibrator for connection to 

this reference probe. Listed below are a few calibration 

points and the resistance readings you can expect from 

the reference probe.

ARTWORK/PRODUCT ART/

 DWGS/START HERE ARROW

5

ARTWORK/PRODUCT ART/

 DWGS/START HERE ARROW

6

BB701

Blackbody Calibrator

omega.com   [email protected]

®

Servicing North America:

U.S.A.:

 

OMEGA Engineering, Inc., One Omega Dr. 

 

P.O. Box 4047, Stamford, CT 06905-0047 USA 

 

 Toll-Free: 1-800-826-6342 (USA & Canada Only) 

Customer Service: 1-800-622-2378 (USA & Canada Only)  

Engineering Service: 1-800-872-9436 (USA & Canada Only) 

Tel: (203) 359-1660 

Fax: (203) 359-7700 

 

e-mail: [email protected]

For Other Locations Visit omega.com/worldwide

It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety 

and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of 

its products to the European New Approach Directives. OMEGA will add the CE  

mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but 

OMEGA accepts no liability for any errors it contains, and reserves the right to 

alter specifications without notice. 

WARNING: These products are not designed for use in, and should not be used  

for, human applications.

WARRANTY/DISCLAIMER

OMEGA ENGINEERING, INC. warrants this unit to be free of defects 

in materials and workmanship for a period of 

13 months

 from 

date of purchase. OMEGA’s WARRANTY adds an additional one (1) 

month grace period to the normal 

one (1) year product warranty

 

to cover handling and shipping time. This ensures that OMEGA’s 

customers receive maximum coverage on each product. 
If the unit malfunctions, it must be returned to the factory for 

evaluation. OMEGA’s Customer Service Department will issue an 

Authorized Return (AR) number immediately upon phone or written 

request. Upon examination by OMEGA, if the unit is found to be 

defective, it will be repaired or replaced at no charge. OMEGA’s 

WARRANTY does not apply to defects resulting from any action of 

the purchaser, including but not limited to mishandling, improper 

interfacing, operation outside of design limits, improper repair, 

or unauthorized modification. This WARRANTY is VOID if the unit 

shows evidence of having been tampered with or shows evidence of 

having been damaged as a result of excessive corrosion; or current, 

heat, moisture or vibration; improper specification; misapplication; 

misuse or other operating conditions outside of OMEGA’s control. 

Components in which wear is not warranted, include but are not 

limited to contact points, fuses, and triacs.

OMEGA is pleased to offer suggestions on the use of its various 

products. However, OMEGA neither assumes responsibility for 

any omissions or errors nor assumes liability for any damages 

that result from the use if its products in accordance with 

information provided by OMEGA, either verbal or written. 

OMEGA warrants only that the parts manufactured by the 

company will be as specified and free of defects. OMEGA 

MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF 

ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT 

THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING 

ANY WARRANTY OF MERCHANTABILITY AND FITNESS 

FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. 

LIMITATION OF LIABILITY: The remedies of purchaser set forth 

herein are exclusive, and the total liability of OMEGA with 

respect to this order, whether based on contract, warranty, 

negligence, indemnification, strict liability or otherwise, 

shall not exceed the purchase price of the component upon 

which liability is based. In no event shall OMEGA be liable for 

consequential, incidental or special damages. 

CONDITIONS: Equipment sold by OMEGA is not intended to be used, 

nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), 

used in or with any nuclear installation or activity; or (2) in medical 

applications or used on humans. Should any Product(s) be used in or 

with any nuclear installation or activity, medical application, used on 

humans, or misused in any way, OMEGA assumes no responsibility 

as set forth in our basic WARRANTY/DISCLAIMER language, and, 

additionally, purchaser will indemnify OMEGA and hold OMEGA 

harmless from any liability or damage whatsoever arising out of the use 

of the Product(s) in such a manner.

RETURN REQUESTS/INQUIRIES

Direct all warranty and repair requests/inquiries to the OMEGA 

Customer Service Department.  BEFORE RETURNING ANY 

PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN 

AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER 

SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING 

DELAYS). The assigned AR number should then be marked on the 

outside of the return package and on any correspondence.
FOR 

WARRANTY

 RETURNS, 

please have the following 

information available BEFORE 

contacting OMEGA:
1. Purchase Order number under 

  which the product was  

 PURCHASED, 

2. Model and serial number of    

  the product under warranty, and 

3. Repair instructions and/or    

  specific problems relative to 

  the product.

FOR 

NON-WARRANTY

 REPAIRS, 

consult OMEGA for current repair 

charges. Have the following 

information available BEFORE 

contacting OMEGA:
1.  Purchase Order number to 

 cover 

  the COST of the repair or  

 calibration,

2.  Model and serial number of the 

  product, and 

3.   Repair instructions and/or specific 

problems relative to the product.

OMEGA’s policy is to make running changes, not model changes, 

whenever an improvement is possible. This affords our customers 

the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© Copyright 2013 OMEGA ENGINEERING, INC. All rights reserved.  

This document may not be copied, photocopied, reproduced, 

translated, or reduced to any electronic medium or machine-readable 

form, in whole or in part, without the prior written consent of OMEGA 

ENGINEERING, INC.

MQS3264/0813

Internal Reference Probe Connections

RS232 Communication

The RS232 communications port allows bi-directional 

data transfer via a three conductor cable consisting 

of signal ground, receive input, and transmit output. 

It is recommended that less than fifty feet of cable be 

used between the computer and this instrument. Use a 

straight DB9 (female) to DB9 (male) connector cable to 

connect your computer to the BB701. The cable should 

be attached only when the computer and BB701 are off.

Internal Reference Probe Connections

 Target Plate Temperature Reference Probe Resistance
 

-10°C (14°F) 

96.09 Ohms

 

+10°C (+50°F) 

103.90 Ohms

 

+50°C (+122°F) 

119.40 Ohms

 

+70°C (+158°F) 

127.07 Ohms

 

+100°C (+212°F) 

138.05 Ohms

 

+120°C (+248°F) 

146.06 Ohms

For complete product manual:  

www.omega.com/manuals/manualpdf/M3264.pdf

230 VAC Model Only

1.  Press the   key to enter the programming mode. 

The lower display will alternately display the menu 

level and “Ac.Cd.”

2.  Use the   and   keys to change to the desired 

menu level.

3.  Once you have chosen the desired menu use the 

key to scroll through the parameters. To change the 

setting of a given parameter, use the 

and 

keys.

4.  To save settings press the   key. The controller 

now exits the programming menu and return to the 

normal operating mode.

5.  To change settings on other menu levels, you must 

re-enter the programming menu (from step #1).

TARGET PLATE ASSEMBLY

RTD REFERENCE PROBE

REFERENCE

PROBE

CONNECTOR

BLACK WIRE

1

2

3

REAR PANEL

RED  WIRE

BLACK WIRE

1

6

7

8

9

6

7

8

9

2

3

4

5

1

2

3

4

5

RX

TX

GND

TX

RX

GND

BB701

BLACKBODY

CALIBRATOR

COMPUTER

SERIAL 

PORT

(RS232)