Page 3
SPECIFICATIONS
(Typical @ 25C and rated supply voltage unless otherwise specified)
STANDARD INPUTS
* 4-20ma 0-50ma 0-5Vdc 0-10Vdc
* Thermocouple inputs: J, K, T, E, R, S, and B
* Cold junction compensation error: +/- 1C (10C to 40C)
* Open input indication:
HELP
displayed
* Temperature displayable in Degrees C or F
* Non-standard inputs available --- consult factory.
ANALOG TO DIGITAL CONVERSION
* 4-1/2 Digit (20,000) Count) A/D Converter
* Dual slope integrating converter with 7 conversions /sec. (typical) rate.
SCAN RATE
* Fixed: two channels per second
DISPLAY
* Red seven segment displays--0.8" (20mm) digit height
* Over range indication:
HELP
* Display test: Briefly displays 8.8.8.8.8.8.8 on power up.
*
Seven-digit
display
SCALE/OFFSET
* Scale programmable from 1 - 30000 units
* Offset programmable from 0 - 99.99mA (For mA Input)
* Offset programmable from 0 - 9.999V (For Voltage Input)
* Programmable decimal point.
RELIABILITY/ACCURACY
*Calibration: NIST traceable (for thermocouples)
*Temperature resolution: 1C/1F
* Warranty: 1 Year
* Recalibration recommended at 12-month interval
POWER REQUIREMENT
* 120Vac --- 60 HZ
(Standard)
* 240Vac --- 50 HZ
(optional)
*
15VDC (optional)
OUTPUT (optional)
* Open collector outputs -- 50ma maximum sink, limited to 50ma per output
* Relays: Single pole single throw, 1 Amp @ 28Vdc or 0.5 Amp @ 120Vac resistive
*Output termination: Euro-style plug-in connector