![Omega Engineering OMEGAFLEX FPU5-MT-110 User Manual Download Page 60](http://html1.mh-extra.com/html/omega-engineering/omegaflex-fpu5-mt-110/omegaflex-fpu5-mt-110_user-manual_3931713060.webp)
Troubleshooting Guide
7-1
7
7.1 Display Error Messages
Error Message
Problem
Solution
CAL OVERFLOW!
While calibrating, you
1.
Repeatedly press the
have allowed the motor
CALIBRATE key
until
to run at a high RPM
the message “Prime or
setting for a very long
START?” appears.
time, or you have entered
a total volume that would
result in an inappropriate
calibration factor.
Calibration FAIL
Your attempt to calibrate
1.
Repeatedly press the
was unsuccessful.
CALIBRATE key
until
the message “Prime or
START?” appears.
2.
Repeat calibration but only
allow the motor to run for a
minute or two.
NOTE: Make sure that you
measure the volume of the
dispensed fluid accurately.
Illegal Op Code
A
A
The line voltage may
1.
Use a multimeter to measure
be bad.
the line voltage used by the
pump motor.
2.
Compare the voltage rating
on the unit’s label (on the back
of the pump motor) with the
line voltage to make sure that
you are using the correct volt-
age. The pump motor
requires line voltages of 90 to
130 Vac (for 115 Vac units)
or 190 to 260 Vac (for 220
Vac units).
2. Repeat calibration but only
allow the pump to run for a
minute or two. N
OTE: Make
sure that you measure the
volume of the dispensed fluid
accurately.
Summary of Contents for OMEGAFLEX FPU5-MT-110
Page 6: ...Peristaltic Pump Motor iv TABLE OF CONTENTS Notes ...
Page 8: ...Index I II Notes ...
Page 9: ...Index I III ...
Page 11: ...Introduction 1 1 2 Notes ...
Page 19: ...Parts of the Pump Motor 2 2 8 Notes ...
Page 25: ...Getting Started 3 3 6 Notes ...
Page 70: ...Troubleshooting Guide 7 7 11 Notes ...
Page 72: ...Technical Details 8 8 2 Notes ...
Page 78: ...Specifications 9 9 6 Notes ...