background image

 

 

11.  Calibration 

When the instrument is first installed , or after a long period of downtime, a moderate zero shift is normal. This is due 
to the rubber diaphragms which are stretched by the internal springs.  After a few operations, the instrument will 
settle into its normal operating condition.  It is recommended that, under these circumstances, instruments should be 
exercised by alternately applying zero and full scale signals several times.  Zero calibration should then be carried 
out. 

Air Supply 

Minimum signal 

Adjust zero control (anti-clockwise)  to 
give minimum required output pressure 

+ to Maximum Signal 

Adjust range control (anti-clockwise on 2-wire, clockwise on 3 
wire)  to give maximum required output pressure 

Note: Reverse Acting Operation 

1 About 20 turns of the zero screw may 

be required to reset the zero point. 

2 3-wire instruments require to be factory 

set for reverse action. 

3 Reverse action is not currently 

available on high pressure models 

4 Do not over wind nozzle to prevent 

jamming 

a

b

c

Reviews: