8-1
Specifications
8
Section 8 - Specifications
Measuring Range:
0 to 100%
Accuracy Range:
±2.5% from 20 to 80% RH, ±3.5% from 5 to 20%
and 80 to 95% RH; ±4% from 0 to 5% and 95 to
100% RH
Hysteresis:
±1%
Repeatability:
±0.1%
Resolution:
0.1%
Response Time:
8 Seconds typical
Sample Rate:
1 sample every 4 seconds.
Operating Temperature:
-30 to 75°C
Input Power
4-20mA:
9 to 30 Vdc @ 50mV
0-1V:
9 to 30 Vdc @ 50mA
0-5V:
9 to 30 Vdc @ 50mA
0-10V:
12 to 30 Vdc @ 50mA
Max Loop Resistance:
200 Ohms @ 9 Vdc supply voltage
(4 to 20mA)
1,300 Ohms @ 30 Vdc supply voltage
Ohms = ((V supply – 4 V) / .02 A)
Max. Load (min. resistance)
Resistance:
1.25K for outputs: 0 – 1 Vdc, 0 – 5 Vdc
2.40K for 0 – 10 Vdc output
Sensor Type:
Digital Sensor
Enclosure Housing:
Gray Polycarbonate, NEMA 13 (IP 54).
Connections:
Nylon, Liquid-tight with neoprene gland for 2.29
to 6.73 m (0.09 to 0.265") diameter cable, internal 5
position terminal block accepts wire size from 14
to 22 gauge wire.
Dimensions:
See “Mounting” Section
Weight:
HX92B (*):
124 g (4.4 oz)
HX92B (*)-D:
132 g (4.7 oz)
HX92B (*)-RP1:
230 g (8.1 oz)
* 4 to 20 mA, 0 to 1 volt, 0 to 5 volts or 0 to 10 volts output. See page 2.2 for
model numbers.