background image

 

 

www.microscopenet.com

 

 

- 6 -

65 

3

 

Operation 

3.1 Adjusting illumination 

1)  Connect the power cord and turn on the main switch.   
2)  Turn the variable intensity dial to increase or decrease the brightness. 

3.2 Placing specimen 

1)  Place the slide on the mechanical stage.   
2)  Use the slide holder to gently secure the slide.   
3)  Turn the X and Y translational knobs to position the specimen for viewing. 

Caution:

 Be sure not to allow an objective to touch a specimen slide when changing 

objectives. 

3.3  Adjusting interpupillary distance 

While observing with both eyes, hold the left and right eyepiece tubes then slowly 
slide the tubes in and out. The interpupillary distance is correct when the left and 
right fields of view converge completely into one 
image. 

3.4  Adjusting eyepiece diopter 

1)  Rotate the 10x objective into position. 
2)  Rotate the diopter ring on the right eyepiece tube 

until its numerical value is the same as your 
interpupillary distance, for example, 65 in the 
right figure. 

3)  Close your left eye and bring the specimen into 

focus following the focusing procedures in 3.5. 

4)  Close your right eye and bring the same 

specimen into clear sharp focus by adjusting the 
diopter ring on left eyepiece tube only. Do not 
use focus knobs at this step. 

5)  Since both sides are adjustable, you may also do 

the above in the opposite way, in other words, left 
eye first and right eye second.   

3.5 Focusing 

1)  With the 10x objective in position, raise the mechanical stage using the coarse focus 

knob until the specimen is close to the objective 

2)  Turn the coarse focus knob until the specimen is in focus.   
3)  Use the fine focus knob to obtain a sharp image.   
4)  You may now switch to another magnification objective. 

Tips: 

To prevent your specimen slide from making contact with an objective, raise the 

stage to its highest position without contacting the 100x objective, then tighten the 
upper limit mechanical stage lever.

 

3.6 Adjusting condenser 

1)  Turn the condenser focus knob to raise or lower the condenser.   

Summary of Contents for M828T

Page 1: ...User Manual Binocular Compound Microscope Model M828T MicroscopeNet com...

Page 2: ...Table of Contents i Caution 3 ii Care and Maintenance 4 1 Component Illustration 5 2 Installation 6 3 Operation 8 4 Specifications 10 5 Troubleshooting Guide 11...

Page 3: ...ve or any other component disconnect the power cord immediately and wipe up the spillage Otherwise the instrument may be damaged 5 Important the lamp lamp housing and adjacent parts will become very h...

Page 4: ...he optical surface use a soft cloth or cotton swab dampened with a lens cleaning solution available at camera stores All optical lenses should be swabbed using a circular motion A small amount of abso...

Page 5: ...tch 3 Eyepiece Tube 11 Head Thumb Screw 19 Condenser Focus Knob 4 Nosepiece 12 Microscope Body 20 Focus Stop Lever 5 Objective 13 Focus Tension Ring 21 Abbe Condenser 6 Slide Holder 14 Coarse Focus Kn...

Page 6: ...hen changing the objective magnification rotate the objective nosepiece until you hear a click sound This ensures the objective is centered in the optical light path 2 4 Installation of filter 1 Swing...

Page 7: ...el may not have one 1 Unplug the power cord 2 Screw off the light cover on the base 3 Screw the black disc onto the base and then insert the mirror into the hole at the center of the black disc You ma...

Page 8: ...e is the same as your interpupillary distance for example 65 in the right figure 3 Close your left eye and bring the specimen into focus following the focusing procedures in 3 5 4 Close your right eye...

Page 9: ...diaphragm is designed to adjust the aperture size not to adjust brightness Generally opening the diaphragm to 70 80 of the N A value of the respective objective will provide an image of acceptable qua...

Page 10: ...ocusing adjustment at 0 002mm range 28mm Stage Mechanical stage Dimension 5 1 2 x 5 1 2 140mmx140mm Movement range 3 x 2 75mm X 50mm Illumination Transmitted 6V 20W Halogen Variable intensity Power Su...

Page 11: ...quality No slide cover attached to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down Turn slide over...

Page 12: ...nsity control dial Aperture diaphragm closed too far Open to the proper setting Condenser position too low Position the condenser upward MECHANICAL PROBLEMS Problem Cause Solution Image will not focus...

Reviews: