3.4  Adjusting eyepiece diopter 

1)  Rotate the 10X objective into position. 
2)  Rotate the diopter rings on the eyepiece tubes 

until its numerical value is the same as your 
interpupillary distance, for example, 

70

 in the 

figure (See 

Fig. 6

). 

3)  Close your left eye and bring the specimen 

into focus following the focusing procedures in 

3.5

4)  Close your right eye and bring the same 

specimen into clear sharp focus by adjusting 
the diopter ring on left eyepiece tube only. Do 
not use focus knobs at this step. 

5)  Since both sides are adjustable, you may also 

do the above in the opposite way, in other 
words, left eye first and right eye second.   

3.5  Focusing 

1)  With the 10X objective in position, raise the mechanical stage using the coarse 

focus knob until the specimen is close to the objective. 

2)  Turn the coarse focus knob until the specimen is in focus.   
3)  Use the fine focus knob to obtain a sharp image.   
4)  To get a good focused image, you may need to combine the focus knob adjustment 

and interpupillary distance adjustment, along with eyepiece diopter adjustment 
stated in 

3.3

 and 

3.4

5)  You may now switch to another magnification objective. 

Tips:   

To prevent your specimen slide from making contact 
with an objective, raise the stage to its highest position 
without contacting the 100X objective, then tighten the 
focus stop lever (

Fig. 7

).

 

Give the stage a tiny extra 

moving space to ensure the objective can be focused 
every time. 

3.6  Applying the immersion oil 

1)  Rotate the objective nosepiece to seat the observing position between the 40X and 

100X objectives as shown in 

Fig. 8

 

(a)

2) 

Place a drop of immersion oil on the slide cover as shown in 

Fig. 8

 

(b)

.

 

3)  Rotate the objective nosepiece to seat the 100X objective to the observing position 

until you hear a “click” sound. 

4)  After observing the specimen, use the camera cleaning kit to clean the 100X 

objective lens gently and the specimen in time. 

5)  If it is hard to clean, you need a little bit alcohol to clean the 100X objective lens and 

the specimen. 

 
 
 

Fig.6 

Fig.7 

70 

Focus Stop Lever 

Summary of Contents for M827TL Series

Page 1: ...User Manual Binocular Compound LED Microscope Model M827TL Series MicroscopeNet com...

Page 2: ...tion 1 ii Care and Maintenance 2 2 Components Illustration 3 3 Installation 4 4 Operation 6 5 Specifications 10 6 Optional Parts 11 7 Troubleshooting Guide 13 8 Darkfield Condenser Installation and Op...

Page 3: ...of direct sunlight high temperature or humidity and dusty environments Ensure that the microscope is located on a smooth level and firm surface 4 If any specimen solutions or other liquids splash onto...

Page 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Page 5: ...Screw 19 Focus Stop Lever 4 Nosepiece 12 Microscope Body 20 Condenser Lock Thumb Knob 5 Objective 13 Focus Tension Ring 21 Condenser Focus Knob 6 Slide Holder 14 Coarse Focus Knob 22 Abbe Condenser 7...

Page 6: ...mechanical stage is at its lowest position 2 Turn the caps counter clockwise to remove them from the nosepiece 3 Take the objectives out from the plastic cases and turn each one clock wise into the h...

Page 7: ...take the eyepieces off and be certain that the head is securely locked by the thumb screw 2 6 Installing the mirror optional may not included in your package 1 Turn off the power switch and disconnec...

Page 8: ...d into the power socket on the microscope and connect it to the power outlet 2 Turn on the power switch 3 Rotate the brightness intensity dial to increase or decrease the brightness Caution A diffusio...

Page 9: ...get a good focused image you may need to combine the focus knob adjustment and interpupillary distance adjustment along with eyepiece diopter adjustment stated in 3 3 and 3 4 5 You may now switch to...

Page 10: ...s first 3 7 Adjusting condenser 1 Turn the condenser focus knob to raise or lower the condenser 2 Raise the condenser when using high power objectives and lower it when using low power objectives Note...

Page 11: ...by following the procedures in 3 5 2 Install the camera by following the procedures in 2 7 3 Open image observing sofeware to examine more details see camera s manual 4 You also can capture images or...

Page 12: ...and filter holder Rack and pinion adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on both sides with focus stop Minimum fine focusing adjustment at 0 002mm range 28mm Mechanical Stag...

Page 13: ...d separately 1 Cameras Model Resolution Operating System Software A1502 640 x 480 0 3MP MS Windows Mac OS X 10 6 10 8 Included A3513U 1280 x 1024 1 3MP MS Windows Mac OS 10 8 and up Linux 2 6 or above...

Page 14: ...ze diameter A191 Dry 0 7 0 9 37mm A191BD Oil 1 36 1 25 37mm A191BOIL Oil 1 36 1 25 Plan 100X 1 25 oil 0 5 160 0 17 spring w iris diaphragm 37mm 3 Carrying Case Model Dimension Net Weight A187V 9 7 8 x...

Page 15: ...is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down specimen at the bottom Turn slide over so the cover glass faces up Diopter adjustment is not set pr...

Page 16: ...e condenser is not in the right position Lower or raise the condenser slightly to the position that the specimen is the brightest A191BOIL A191BD only Make sure the oil contact the bottom of the slide...

Page 17: ...ondenser up or down slightly to get the best darkfield viewing Note The dry darkfield condenser is used with the dry objectives only The dry darkfield condenser works with the 4X 10X 40X objectives Th...

Page 18: ...jective you will need to drop the immersion oil and adjust the iris ring Fig 15 to get proper brightness and contrast of view field 8 After observing the specimen with 100X objective use the lens clea...

Reviews: