- 15 - 

7.

 

Darkfield Condenser Installation and Operation 
Instructions 

7.1  Dry darkfield condenser of A191 

1)  Mounting the dry darkfield condenser 

  Loosen the condenser lock thumb screw on the 

condenser holder and remove the brightfield 
condenser. 

  Install the dry darkfield condenser and tighten 

the condenser lock thumb screw on the 
condenser holder (see 

Fig. 12

)

2)  Centering the dry darkfield condenser 

  Turn the 4X objective to the light path.   

  Turn the condenser focus knob to lower the condenser till a dark spot showed in 

the viewing field as shown in 

Fig. 13

 

(a)

  Turn the condenser translational centering screws to move the dark spot to the 

center as shown in 

Fig. 13

 

(b)

3)  Place the slide on the stage.   
4)  Raise the condenser all the way to the top and 

lower it a little bit.   

5)  Following the procedures in this manual to focus 

and observe. 

6)  Move the condenser up or down slightly to get the 

best darkfield viewing. 

Note

 

  The dry darkfield condenser is used with the dry objectives only. 

  The dry darkfield condenser works with the 4X, 10X, 40X objectives. 

 

The dry darkfield condenser won

’t work with the 100X oil immersion objective.

 

 

7.2  Oil darkfield condenser of A191BOIL 

1)  Mounting the oil darkfield condensers 

  Loosen the condenser lock thumb screw on 

the condenser holder and remove the 
brightfield condenser. 

  Install the oil darkfield condensers and tighten 

the condenser lock thumb screw on the 
condenser holder (see 

Fig. 14

)

2)  Centering the oil darkfield condensers 

  Turn the 40X objective to the light path.   

  Turn the condenser focus knob to slowly lower 

and raise the condenser till a dark spot showed in 
the viewing field as shown in 

Fig. 15

 

(a)

  Turn the condenser translational centering 

screws to move the dark spot to the center as 
shown in 

Fig. 15 (b)

.

 

Condenser Lock   

Thumb Screw 

Centering Screws 

Condenser Holder 

 

     

(a)                                (b) 

 

   

 

   

(a)                                (b) 

Fig.12 

Fig.13 

Fig.14 

Fig.15 

Condenser Lock   

Thumb Screw 

Centering Screws 

Condenser Holder 

Summary of Contents for M824 Series

Page 1: ...User Manual Trinocular Compound Microscope Model M834 Series MicroscopeNet com...

Page 2: ...1 Component Illustration 3 2 Installation 4 3 Operation 7 4 Specifications 10 5 Optional Parts 11 6 Troubleshooting Guide 13 7 Darkfield Condenser Installation and Operation Instructions 15 8 Phase C...

Page 3: ...l electrical connectors power cord should be inserted into an electrical surge suppressor to prevent damage due to voltage fluctuations 6 Important the lamp lamphouse and adjacent parts will become ve...

Page 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Page 5: ...11 Microscope Base 19 Condenser Holder 4 Slide Holder 12 Eyepiece 20 X Y Stage Moving Knobs 5 Condenser Lock Thumb Screw 13 Diopter Adjusting Ring 21 Light Collector 6 Condenser Focus Knob 14 Eyepiec...

Page 6: ...iece tubes 2 Insert the eyepieces into the eyepiece tubes 2 4 Installation of the objectives 1 Adjust the coarse focus knob until the mechanical stage is at its lowest position 2 Turn the caps counter...

Page 7: ...ight housing 3 Take the eyepieces off and check if the view head is securely locked by the set screw Put the microscope down on its back gently 4 Open the cover on the bottom of the base by pull the k...

Page 8: ...oto tube and then connect the camera to a computer via USB2 0 cable See Fig 6 3 The manual for the camera is on a CD or mini CD Refer to the manual to install the driver and software on to the compute...

Page 9: ...il the specimen is close to the objective 2 Turn the coarse focus knob until the specimen is in focus 3 Use the fine focus knob to obtain a sharp image 4 To get a good focused image you may need to co...

Page 10: ...sting condenser 1 Turn the condenser focus knob to raise or lower the condenser 2 The condenser is raised when using high magnification objectives and lowered when using low magnification objectives N...

Page 11: ...microscope into focus by following the procedures in 3 3 2 Install the camera by following the procedures in 2 8 3 Open image observing software to examine more details see camera s manual 4 If the im...

Page 12: ...Coaxial coarse and fine focusing knobs on both sides w stage upward stopper Minimum fine focusing adjustment at 0 002mm range 30mm Mechanical Stage Large double layer mechanical stage Dimension 5 1 2...

Page 13: ...S Windows 32 64 bit Mac OS Included A1510 1280 x 1024 1 3MP A1520C 1600 x 1200 2 0MP A1530X 2048 x 1536 3 0MP A1550X 2592 x 1944 5 0MP A1590 3488 x 2616 9 0MP MS Windows 32 64 bit 2 Darkfield Condense...

Page 14: ...Achromatic 20X with built in phase plate Achromatic 40X with built in phase plate spring Achromatic 100X with built in phase plate spring oil five positions 10 for 10X phase contrast objective 20 for...

Page 15: ...17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down specimen at the bottom Turn slide over so the cover glass faces up Imm...

Page 16: ...e condenser is not in the right position Lower or raise the condenser slightly to the position that the specimen is the brightest A191BOIL only Make sure the oil contact the bottom of the slide and th...

Page 17: ...e condenser up or down slightly to get the best darkfield viewing Note The dry darkfield condenser is used with the dry objectives only The dry darkfield condenser works with the 4X 10X 40X objectives...

Page 18: ...objectives a Take off all the objectives from the nosepiece b Install the phase contrast objectives onto the nosepiece following the steps in 2 4 2 Mounting the phase contrast condenser annular ring d...

Page 19: ...on focusing etc following the instructions in this manual Note The phase contrast condenser will be working as a conventional Abbe condenser if the annular ring disk being put at B position 8 2 Phase...

Page 20: ...he telescope The bright ring and dark ring should be coincided with each other as shown in Fig 21 d e If the ring images are not clear turn the top of telescope until both ring images are in focus f I...

Page 21: ...is hard to find put a normal white print paper on the stage and under the objective then you will see the dark ring as shown in Fig 24 a h If the bright ring is still obscure as in Fig 24 b raise or l...

Reviews: