PV8355F4E007 15
(13) Confirm
the
results.
→
See
“
4.14 Checking the results
.
”
(14) Shut
down
the
software
and
then
shut
down the controller supply unit and computer.
→
See
“
4.15 Termination
.
”
4.2. Start-up
(1) Set the power switch of the controller supply unit to ON.
(2) Set the lamp 2 switch of the controller supply unit to ON.
(3) Start the computer, double-click the shortcut icon on the desktop, and start the
program.
4.3. Revolver settings
(1) Set a revolver.
A revolver is adjusted to the through position (in which an objective lens is not
installed) when measuring a spectral transmission factor using a transmissit unit.
㻌
㻌
㻌
㻌
㻌
㻌
㻌
㻌
㻌
For the handling of the computer, see the computer
’
s instruction
manual for correct usage procedure.
It takes about 15 minutes to set the lamp switch to ON and to
stabilize a halogen lamp after it lights. Warm up the halogen lamp
as required.
Revolving nosepiece