UHI-4 Insufflator Quick Reference Guide
Insufflator Controls
Cavity Modes
Flow Rates
There are two cavity modes Normal and Small which regulate the pressure and flow rates.
The are three flow settings:
Note:
The cavity mode cannot be changed once insufflation has started, if required stop
insufflation and then change mode.
Warning:
when using a Veress needle always use the LOW flow mode as insufflation in other
modes is dangerous because the cavity pressure may exceed the set pressure.
Press to toggle between normal and small
Cavity mode
Setting pressure range
Note
NORMAL
3 - 25 mmHg
(0.4 - 3.3 kPa)
To increase the pressure from 20
to 21 mmHg, keep pressing “+”
for at least three seconds
SMALL
3 - 15 mmHg
(0.4 - 2.0 kPa)
Cavity mode
Setting pressure range
Note
NORMAL
HIGH
MED
LOW
20 - 45
1.1 - 1.9, 2 - 5
0.1 - 1.0
SMALL
HIGH
MED
LOW
6 - 10
1.1 - 1.9, 2 - 5
0.1 - 1.0
Press to toggle between flow rates
Smoke Evacuation
Connector
Power Switch
Set Pressure
Flow Rate
Actual Pressure
Volume Indicator
CO
2
Connector
Smoke Evacuation
Speed
Start/Stop