UHI-4 Insufflator Quick Reference Guide

Insufflator Controls

Cavity Modes

Flow Rates

There are two cavity modes Normal and Small which regulate the pressure and flow rates.

The are three flow settings:

Note:

 The cavity mode cannot be changed once insufflation has started, if required stop 

insufflation and then change mode.

Warning: 

when using a Veress needle always use the LOW flow mode as insufflation in other 

modes is dangerous because the cavity pressure may exceed the set pressure.

Press to toggle between normal and small

Cavity mode

Setting pressure range

Note

NORMAL

3 - 25 mmHg  

(0.4 - 3.3 kPa)

To increase the pressure from 20  
to 21 mmHg, keep pressing “+”  
for at least three seconds

SMALL

3 - 15 mmHg 

(0.4 - 2.0 kPa)

Cavity mode

Setting pressure range

Note

NORMAL

HIGH

MED
LOW

20 - 45

1.1 - 1.9, 2 - 5

0.1 - 1.0

SMALL

HIGH

MED
LOW

6 - 10

1.1 - 1.9, 2 - 5

0.1 - 1.0

Press to toggle between flow rates

Smoke Evacuation

Connector

Power Switch

Set Pressure

Flow Rate

Actual Pressure

Volume Indicator

CO

2

 Connector

Smoke Evacuation

Speed

Start/Stop

 

Issue 1 

05/02/2021

 

KMF-08929 

NCDR 02867