Combining this product with the microscope frame (BX43F, BX53F2) and the control box for coded function (U-CBS) makes
it possible to use the light manager function of the microscope frame and the function to readout the revolving nosepiece
information to the digital camera/imaging software at the same time.
AX8099 05
Optical Microscope Accessory
U-IFRES
INSTRUCTIONS
INTERFACE FOR CODED NOSEPIECE
1) For operating procedures of the microscope system attached with this product, refer to the instruction manuals
provided with the microscope frame and U-CBS.
2) Before connecting the cables, make sure that the powers of the microscope frame and U-CBS are set to OFF.
3) Do not stretch or bend the cables with the excess force. Otherwise, they may be damaged. When connecting the
cables, be sure to lay out the cables according to the instruction manual provided with U-CBS.
4) When carrying the microscope system, be careful so that the cables are not pinched under the microscope frame.
5) Before disposing of this product, be sure to check the regulations and rules of your local government and follow
them. Contact Olympus for any questions.
1
Cautions
2
Attaching the body of interface for coded nosepiece (U-IFRES)
Clean the area to attach the fixing board on the back side of the
microscope frame with a soft cloth.
Remove the protection sticker from the fixing board.
Attach the fixing board to the back side of the microscope frame firmly.
If you remove the sticker of the fixing board attached once, the
adhesive power may be weakened and U-IFRES may drop.
Decide the position carefully before attaching to avoid the
double attachment.
When using BX53F2
Align the fixing board with the upper left of the back side of the
microscope frame.
1
2
3
TIP
Connect the connector of the coded revolving nosepiece to U-IFRES.
· Insert the connector by matching marks a .
· For coded revolving nosepiece, refer to the instruction manual
for U-D7RES/U-D6RES.
4
TIP
Attach U-IFRES by hooking the pin b of U-IFRES on the groove c
of the fixing board.
If you attach or remove the connector while U-IFRES is being
hooked on the fixing board, the sticker of the fixing board may
come off and U-IFRES may drop. When you attach or remove
the connector, be sure to hold U-IFRES with your hands.
5
TIP
a
4
3
Layout of the cable
4
Specifications
Refer to the instruction manual provided with the U-CBS.
Item
Specifications
U-IFRES
Dimension
Body: 44 (W) x 25 (D) x 60 (H) mm
Fixing board: 40 (W) x 19 (D) x 38 (H) mm
Weight
0.15 kg
Operating
environment
· Indoor use
· Altitude: Max. 2,000 meters
· Ambient temperature: 5 to 40 °C (41 to 104 °F)
· Maximum relative humidity: 80% for temperatures up to 31 °C (88 °F) (without condensation)
In case of over 31 °C (88 °F), the relative humidity is decreased linearly through 70% at 34 °C (93 °F),
60% at 37 °C (99 °F), and to 50% at 40 °C (104 °F).
· Supply voltage fluctuations: Not to exceed ±10% of the normal voltage.
· Pollution degree 2 (in accordance with IEC60664-1)
· Installation/Overvoltage category: II (in accordance with IEC60664-1)
b
c
5
BX53F2
b
c
5
BX43F
If the abnormality or defect is found with the product, contact Olympus for assistance. At that time, please tell them
the following information as well.
· Product name and abbreviation (Example: interface for coded nosepiece U-IFRES)
· Serial No. (SN)
· Problem
Repair request
When using BX43F
Align the fixing board with the upper right corner of the label on the back
side of the microscope frame.
Be careful not to overlap with the label.
TIP
1
2
3
BX43F
This product is a part of the system product which complies with the CE marking.
Please refer to the instruction manual of your system product for the safety instructions related to the CE marking.
1
2
3
BX53F2